2022/12/10 (Saturday)

Previous Day

  • Lab in-personnel:
Status Run 76 ending
Setup DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe removed from lab.
NaI scan in progress
Warming up. Mixture in dump. Mixture removed from LN2 Trap 1 but it still has contaminants. Some He exchange gas in cryostat vacuum jacket and IVC.
1432 Recovery = 700946.39 cuft (@ 14:32:41), 0.08 cuft/hr, much lower. Time to close the recovery valve.
1450 MC(Cernox)/Sorb = 168/>250 K
1450 MC(Cernox) = 167.53 K, 4.2 K/hr. Implies warming tau of 30.3 hours, should reach dew point (around 34 F) at Dec 12 23:33
2425 MC(Cernox) = 199.34 K, 2.9 K/hr. Implies warming tau of 32.4 hours, should reach dew point (around 34 F) at Dec 13 03:38
2428 Recovery = 700946.39 cuft (@ 04:28:53), 0.05 cuft/hr