2022/12/01 (Thursday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array undergoing scan, trigger threshold varies.
Cooled to base two days ago after twenty-eighth 5K pumpout. Taking SiHV-triggered data overnight
0835 DCRC1pA baseline = -0.795 V
1120 DCRC1pA baseline = -0.787 V. Looks like dump sizes are smaller.
Around 8:00 the MC baseline rose a bit, usually a precursor to losing the phonon channels. At about the same time the dump sizes fell slightly. But no sign of problems in the traces yet. Around 9:45 the dumps started getting a bit smaller. Then around 11:20 the dump sizes became unstable and started falling precipitously after that. So that's probably when we clearly lost the phonons. About 45 hours after cooling to base.
1215 DCRC1pA baseline = -0.737 V. Few pulses, lots of wiggles. Phonon channels are kaput.
1215 Close V6 to stop circulation.
1220 V6 is now fully closed (light is off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1220 Close 1K Pot needle valve
1220 Turn on Still and MC heaters to 20 mW each
1220 Open V3, releasing pressure from Condenser side to Still side
1225 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1310 MC/1KPot = 3.80/3.73 K. G1(fridge) = 710 mbar.
1310 Open V9 to release pressure into the dump.
1310 Stop DAQ programs and unplug DCRCs 1&3
1450 MC = 6.21 K and flat. 1K Pot = 5.9 K and rising (was flat at 4.4 for a while).
1450 Check pressure in IVC: 86 nbar
1455 Start pumping out IVC with turbo pump.
1455 Fill LN2 cold trap dewar
1455 MB = 29.0%
1625 Tower = 10.06 K. Turn off MC and Still heaters

Next Day