start › cdms › k100 › run_diary › 20221122

2022/11/22 (Tuesday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled down to base two days ago after 25th 5K pumpout. Taking 0V data overnight, SiHV-triggered with low threshold.
0950 DCRC1pA baseline = -0.798 V
1045 MC base = 32.0 mK. Most spikes peak below 36 mK, some above 40 mK
1045 DCRC1pA baseline = -0.798 V
1130 DCRC1pA baseline = -0.790 V
1155 DCRC1pA baseline = -0.765 V. Pulses are weak and there are occasional baseline wiggles. He films have killed the phonon channels.
Looking at dump sizes there was a change at 8:10 am. That would be 43 hours after cooling to base - that's close to the amount of time before we lose the phonon channels in recent cycles. And yet we had apparently good baselines later than that. Could be that the variations in baseline come and go, and the problems started around 8:10. There was also an excursion in MC temperature starting around 8:20
1205 Close V6 to stop circulation.
1220 V6 is now fully closed (light is presumably off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1329 MB = 29.5%
1330 Fill LN2 cold trap bath
1334 Turn off DAQ programs and unplug DCRCs1&3
1335 Open V3 to let mixture from Condenser side into Still side.
1336 Close 1K Pot needle valve
1338 MC = 0.64 K
1340 Turn on Still and MC heaters to 20 mW each
1346 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1432 MC/1KPot = 3.88/3.84 K. G1(fridge) = 760 mbar.
1432 Open V9 to release pressure into the dump.
1532 MC/1KPot = 5.87/4.37 K
1535 Check pressure in IVC: 86 nbar
1537 Start pumping out IVC with turbo pump.
1720 Tower = 10.00 K. Turn off MC and Still heaters

Next Day