2022/11/11 (Friday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI scan in progress, not set up for coincidence data
Warmed up yesterday for twenty-second 5K pumpout. Turbo pumping IVC overnight.
1133 Begin LHe transfer @ MB = 9.4%
1140 Fill LN2 cold trap bath
1145 Turbo on IVC = 0.86 nbar. With Turbo valved out for 1 minute IVC = 6.3 nbar.
1145 Close IVC valves at top of fridge.
1150 MB = 37.9%
1150 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.39 bar on mechanical gauge.
1150 Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 gets below 30 mbar open valve all the way.
1155 4He pump valve open all the way.
1155 1K Pot = 3.12 K / 4.0 mbar. G1 (fridge+dump)=419 mbar. MC = 5.70 K.
1210 1K Pot = 2.99 K / 4.0 mbar. G1 (fridge+dump)=372 mbar. MC = 3.26 K.
1223 End LHe transfer @ MB = 94.6 %
1250 1K Pot = 2.52 K / 4.0 mbar. G1 (fridge+dump)=277 mbar. MC = 3.00 K.
1250 Open 1K Pot needle valve → 15°. 1K Pot → 2.35 K / 6.0 mbar
1255 Open 1K Pot needle valve → 30°. 1K Pot → 2.21 K / 7.0 mbar
1300 Open 1K Pot needle valve → 45°. 1K Pot → 2.08 K / 9.5 mbar
1340 MC/Still/1KPot = 2.14/1.95/1.52. G1(fridge+dump) = 59 mbar and getting flat.
1340 Close V3 and V9, leaving 59 mbar mixture in dump (G2)
1340 Turn on 3He pump. G1(condenser) → 74 mbar
1340 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 344 mbar. MC(H35) → 1.64 K. Still → 1.64 K / 80 mbar
1345 V6 → 13.0%. Condenser(G1) peaks at 332 mbar. MC(H35) → 1.44 K. Still(P1) → 1.49 K / 14 mbar.
1350 V6 → 13.4%. Condenser(G1) peaks at 232 mbar. MC(H35) → 1.24 K. Still(P1) → 1.20 K / 4.5 mbar.
1355 V6 → 13.8%. Condenser(G1) doesn't really peak. MC(H35) → 1.07 K. Still(P1) → 1.07 K / 1.4 mbar.
MC wiggled between 1.05 and 1.25 K, Al transition
1410 V6 → 14.4%. Condenser(G1) peaks at 138 mbar. MC(H35) → 0.43 K. Still(P1) → 1.07 K / 0.7 mbar.
1415 V6 → 15.0%. Condenser(G1) peaks at 137 mbar. MC(H35) → 0.30 K. Still(P1) → 1.07 K / 0.4 mbar.
1420 V6 → 100.%. Condenser(G1) peaks at 137 mbar. MC(H35) → 0.15 K. Still(P1) → 1.03 K / 0.095 mbar.
1425 Open V14 to pull the rest of the mixture out of the dump into circulation. Close and reopen 2 times to keep Still pressure below 120 mbar. On last opening Still peaks at 75 mbar, Condenser peaks at 212 mbar, MC peaks at 145 mK, then all relax again. 1K → 1.56 K / 11 mbar.
1450 Still = 0.85 mbar. Close V14
1515 MC = 29.4 mK
1515 1K Pot = 1.52 K / 6.5 mbar, pressure flat. Should open it further.
1515 Open 1K Pot needle valve → 60°. 1K Pot → 2.08 K / 11.5±0.5 mbar
1545 MC = 27.8 mK
1635 MC = 27.4 mK
1635 Plug in DCRCs 1&3
1640 Check phonon channels. Relock and adjust offsets to get all baselines ~ -0.8 V
2040 MC = 29.1 mK
2040 Adjust offsets to get all phonon baselines near -0.800 V
2051 Dial HV up to -108.1V nominal, -106.1 V actual, for 15% prebias (using RootSequencer)
2057 Dial HV down to -94V nominal, -91.8V actual
2100 HV = -91.8 V, Vtest = -21.9 mV, Ileak = -24.1 uA, Vdet = -82.2 V (90%).
2110 Set NaI array trigger window to 4.73 - 5.93 MeV
2113 (meant to start data series but forgot to enable the logger)
2150 HV = -91.8 V, Vtest = -21.9 mV, Ileak = -24.1 uA, Vdet = -82.2 V (90%).
2153 Start -82.2 V PuBe data series 55 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2249 Start -82.2 V PuBe data series 634 good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
2255 DCRC1pA baseline = -0.785 V