2022/11/05 (Saturday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI scan in progress, not set up for coincidence data
Warmed up two days ago for twentieth 5K pumpout. Left IVC closed overnight with all mixture in dump.
1230 Check pressure in IVC: 10 nbar. Close IVC valves again.
1235 MC/1KPot/Sorb = 5.60/>7/13.7 K. G1(condenser)=0.9 mbar, P1(still)=0.14 mbar, G2(dump+Trap1)=675 mbar.
1235 Fill LN2 cold trap bath
1235 Open V12a+speedivalve to let mixture back into fridge.
MC rises from 5.6→6.1 K
1240 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.18 bar on mechanical gauge.
1240 Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 gets below 20 mbar open valve all the way.
1245 4He pump valve open all the way.
1245 1K Pot = 3.31 K / 2.4 mbar. G1 (fridge+dump)=533 mbar. MC =6.04 K.
1300 1K Pot = 3.09 K / 2.4 mbar. G1 (fridge+dump)=490 mbar. MC =5.60 K.
1300 Open 1K Pot needle valve → 15°. 1K Pot → 2.94 K / 5.0 mbar
1305 Open 1K Pot needle valve → 30°. 1K Pot → 2.69 K / 6.5 mbar
MC drops quickly from 5.2 → 3.4 K
1310 Open 1K Pot needle valve → 45°. 1K Pot → 2.69 K / 9.5 mbar
1315 MB = 69.2%
1315 MC/Still/1KPot = 3.22/3.10/2.60. G1(fridge+dump) = 383 mbar and falling.
1335 MC/Still/1KPot = 2.90/2.82/2.38. G1(fridge+dump) = 275 mbar and falling.
MC temperature kinks a couple of times between 2.10 K and 1.97 K, lambda point apparently
1530 MC/Still/1KPot = 1.69/1.74/1.54. G1(fridge+dump) = 33 mbar and pretty flat.
1535 Close V3 and V9, leaving 52 mbar mixture in dump (G2)
1535 Turn on 3He pump. G1(condenser) → 40 mbar
1535 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 226 mbar. MC(H35) → 1.39 K. Still → 1.43 K / 24 mbar
1540 V6 → 13.0%. Condenser(G1) peaks at 230 mbar. MC(H35) → 1.16 K. Still(P1) → 1.19 K / 4.5 mbar.
1545 1K Pot = 1.58 K / 12±1 mbar.
MC temperature kinks around 1.1 K, Al transition apparently
1545 V6 → 13.4%. Condenser(G1) peaks at 175 mbar. MC(H35) → 0.87 K. Still(P1) → 1.13 K / 2.6 mbar.
1555 V6 → 13.8%. Condenser(G1) peaks at 165 mbar. MC(H35) → 0.55 K. Still(P1) → 1.14 K / 1.7 mbar.
1600 V6 → 14.4%. Condenser(G1) peaks at 170 mbar. MC(H35) → 0.24 K. Still(P1) → 1.16 K / 0.7 mbar.
1610 V6 → 15.0%. Condenser(G1) peaks at 160 mbar. MC(H35) → 0.20 K. Still(P1) → 1.14 K / 0.5 mbar.
1615 V6 → 100.%. Condenser(G1) peaks at 164 mbar. MC(H35) → 54 mK. Still(P1) → 1.10 K / 0.13 mbar.
1630 MC = 50.0 mK. Still = 1.10 K / 0.13 mbar. G1/G2 = 150/168 mbar. 1K Pot = 1.58 K / 9.5±0.5 mbar.
1630 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 26 mbar, Condenser peaks at 185 mbar, MC peaks at 69 mK, then all relax again. 1K Pot mostly unaffected.
1725 Still = 0.30 mbar. Close V14
1745 MC = 27.6 mK