2022/11/03 (Thursday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI scan in progress, not set up for coincidence data
Cooled to base two days ago after nineteenth 5K pumpout. As a test we waited a day at 5K with the IVC closed before cooling.
Taking 0V SiHV-triggered data overnight, with DCRC4 disabled (no NaI). Main purpose is just to see when He films kill the phonon channels.
0950 Baselines for DCRC1pA/B/D/DCRC3pA/B/D = -0.734/-0.699/-0.725/-0.702/-0.700/-0.702 V. These are definitely elevated. And pulses are weak.
At 5:30 am the base MC temperature rose from about 30 to about 31 mK, also a familiar sign that He films are about to kill the phonon channels.
At 8:35 am the dump sizes started to change. Rather than dropping off, the increased from a pretty constant 68M to around 100M and became much less stable, falling as low as 45M sometimes.
1030 Stop data series
1030 Close V6 to stop circulation.
1036 V6 is probably now fully closed (light is probably off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1255 Close 1K Pot needle valve
1255 Open V3 to let mixture from Condenser side into Still side.
1255 Turn off DAQ programs and unplug DCRCs 1&3
1255 MC = 0.65 K
1255 Turn on Still and MC heaters to 20 mW each
1300 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1555 MC/1KPot = 5.59/5.91 K. G1(fridge) = 930 mbar. All flat.
1555 Open V9 to release pressure into the dump. (Should have done this earlier)
1705 MC/1KPot = 5.58/5.98 K. G1(fridge+dump) = 547 mbar.
1705 Check pressure in IVC: 0.14 ubar. Start pumping out IVC with turbo pump.
2040 Tower = 3.6 K. Turn off MC and Still heaters. Supposed to do this at 10.0 K, forgot. Hopefully it won't be too hard to cool down.

Next Day