2022/10/27 (Thursday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger thresholds: 7.0 - 11.0 MeV
Cooled to base two days ago after seventeenth 5K pumpout.
Taking -80V NaI-triggered data overnight.
0915 DCRC1pA baseline = -0.792 V
0915 Start -82.0 V PuBe data series 44 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
0915 HV = -91.7 V, Vtest = -21.9 mV, Ileak = -24.1 uA, Vdet = -82 V (89%).
0920 MC base = 30.0 mK. Most spikes peaking below 33 mK, some above 35 mK
0959 Start -82.0 V PuBe data series 366 min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
1325 HV = -91.7 V, Vtest = -22 mV, Ileak = -24.2 uA, Vdet = -82.1 V (89%).
1330 DCRC1pA baseline = -0.794 V
1354 MC/Coldplate = 30.2/151 mK, Still = 1.12 K / 0.065 mbar / 0 mW, G1/G2 = 125/113 mbar, 1K Pot = 45° / 1.58 K / 10±1 mbar, MB = 26.5%
MB relatively low, will have to do an early transfer tomorrow, maybe 9am
1354 Fill LN2 cold trap dewar
1609 Dial HV down to 0V
1611 Close V6 to stop circulation.
1616 V6 is now fully closed (light is off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1617 Close 1K Pot needle valve
1618 Open V3 to let mixture from Condenser side into Still side.
1622 MC = 0.42 K
1622 Turn on Still and MC heaters to 20 mW each
1623 Turn off DAQ programs and unplug DCRCs 1&3
1631 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1650 MC/1KPot = 3.20/3.07 K
1655 Check pressure in IVC: 0.012 ubar, it reads 1.2e-5mbar, it must be low because of the new fast warmup procedure? Start pumping out IVC with turbo pump
1725 MC/1KPot = 3.93/3.94 K. G1(fridge) = 805 mbar
1735 G1(fridge) = 840 mbar. Open V9 to release pressure into the dump.
1755 MC temp start rising more quickly
1937 Tower = 10.01 K. Turn off MC and Still heaters.