2022/10/25 (Tuesday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source in storage in PAN 434.
NaI scan data running overnight
Warmed up last night for seventeenth 5K pumpout.
1048 Turbo on IVC = 1.0 nbar. With Turbo valved out for 1 minute IVC = 10 nbar.
1049 Close IVC valves on top of fridge
1056 Begin LHe transfer @ MB = 1.1%
1100 Fill LN2 cold trap dewar
1111 MB = 21.7%
1112 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.33 bar on mechanical gauge.
1116 Pump out 1K Pot in stages, follow Matt's new procedure of throttling the flow at the valve at the pump itself, opening just until the pump got louder, then waiting, then opening a little more, etc.
1119 Start pumping 1K Pot continuously
1120 MB = 55.4%
1120 MC = 5.83 K, cooling slowly. 1K Pot = 3.20 K / 2.2 mbar
1121 Open 1K Pot needle valve → 15°. 1K Pot → 2.96 K / 6.0 mbar
1125 Open 1K Pot needle valve → 30°. 1K Pot → 2.91 K / 7.0 mbar
1129 Open 1K Pot needle valve → 45°. 1K Pot → 2.82 K / 9.0 mbar
1132 G1/G2 = 364/346 mbar dropping linearly
1132 MC/1Kpot = 3.24/2.78K dropping linearly
1145 Ended LHe transfer @ MB = 91.4%
Forgot to use the heat gun for this transfer
1247 G1(fridge) = 58 mbar. MC = 2.11 K. 1K Pot = 1.52 K / 7.0 mbar
1248 Close V3 and V9, leaving 61.2 mbar mixture in dump
1250 Turn on 3He pump. G1(condenser) → 69 mbar
1250 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 338 mbar. MC(H35) → 1.54 K. Still → 1.58 K / 60 mbar
1300 V6 → 13.0%. Condenser(G1) peaks at 177 mbar. MC(H35) → 1.21 K. Still(P1) → 1.24 K / 5.5 mbar.
1305 V6 → 13.4%. Condenser(G1) peaks at 157 mbar. MC(H35) → 1.05 K. Still(P1) → 1.12 K / 2.8 mbar.
1310 V6 → 13.8%. Condenser(G1) peaks at 156 mbar. MC(H35) → 0.91 K. Still(P1) → 1.07 K / 1.5 mbar.
1315 V6 → 14.4%. Condenser(G1) peaks at 159 mbar. MC(H35) → 0.74 K. Still(P1) → 1.05 K / 0.9 mbar.
1320 V6 → 15.0%. Condenser(G1) peaks at 158 mbar. MC(H35) → 0.53 K. Still(P1) → 1.06 K / 0.60 mbar.
1327 V6 → 100.%. Condenser(G1) peaks at 156 mbar. MC(H35) → 0.21 K. Still(P1) → 1.03 K / 0.16 mbar.
1341 Plug in DCRCs 1 and 3
1345 MC = 50 mK. Still = 1.07 K / 0.10 mbar. G1/G2 = 118/121 mbar. 1K Pot = 1.52 K / 7.0±0.5 mbar.
1346 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at over 30 mbar, Condenser peaks at 204 mbar, MC peaks at 102 mK, then all relax again. 1K Pot unaffected.
1410 Still = 0.75 mbar. Close V14
1445 Bring PuBe source back into lab.
After moving the LN2 dewar the level meter settled at the bottom, it will need to be refilled tomorrow
1450 MC = 28.8 mK. Still = 1.10 K / 0.065 mbar. G1/G2 = 114/117 mbar. 1K Pot = 1.56 K / 10±1 mbar.
1735 Turn on DAQ programs and adjust phonon offsets to get all baselines near -0.80 V
1742 Dial HV up to -108.1V nominal, -106.0 V actual, for 15% prebias (using RootSequencer)
1745 Set NaI array trigger window to 7.0 - 11.0 MeV
1749 Dial HV down to -94V nominal, -91.7V actual
1751 Start -82.1 V PuBe data series 120 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
1755 HV = -91.7 V, Vtest = -21.7 mV, Ileak = -23.9 uA, Vdet = -82.1 V (90%).
1755 DCRC1pA baseline = -0.79 V
1755 MC = 29.4 mK
1956 Start -82.1 V PuBe data series 621 good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
2000 MC = 29.6 mK
2400 DCRC1pA baseline = -0.784 V