UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20221019

2022/10/19 (Wednesday)

Previous Day

  • Lab in-personnel: Zach, Himangshu
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.0 MeV
Warmed up and pumped out IVC overnight for fifteenth 5K pumpout.
1100 Turbo on IVC = 0.86 nbar. With Turbo valved out for 1 minute IVC = 10 nbar.
1101 Close IVC valves on top of fridge
1105 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed).Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar on mechanical gauge.
1110 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1118 Start pumping 1K Pot continuously
1130 MC = 5.45 K, cooling slowly. 1K Pot = 3.31 K / 2.6 mbar, cooling slowly
1130 Open 1K Pot needle valve → 15°. 1K Pot → 3.06 K / 6.5 mbar
1135 Open 1K Pot needle valve → 30°. 1K Pot → 2.96 K / 7.5 mbar
1143 Open 1K Pot needle valve → 45°. 1K Pot → 2.60 K / 11 mbar
1147 G1/G2 = 371/340 mbar dropping linearly
1147 MC/1Kpot = 3.24/2.53K dropping linearly
1325 MC = 1.91 K. It flattened around 2.5 K for about a half hour then dropped quickly to 1.95 K (helium lambda point)
1359 Close V3 and V9, leaving 33.4 mbar mixture in dump
1400 Turn on 3He pump
1406 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 251 mbar. MC(H35) → 1.41 K. Still → 1.45 K / 36 mbar.
1414 Set NaI trigger window to 7-11 MeV
1414 V6 → 13.0%. Condenser(G1) peaks at 275 mbar. MC(H35) → 1.28 K. Still(P1) → 1.57 K / 8 mbar.
1439 V6 → 13.4%. Condenser(G1) peaks at 273 mbar. MC(H35) → 1.14 K. Still(P1) → 1.17 K / 4.0 mbar.
1450 V6 → 13.8%. Condenser(G1) didn't really peak, a small kink in the falling curve. MC(H35) → 0.89 K. Still(P1) → 1.13 K / 1.08 mbar.
1454 V6 → 14.4%. Condenser(G1) peaks at 187.8 mbar. MC(H35) → 0.52 K. Still(P1) → 1.12 K / 1.0 mbar.
1504 V6 → 15.0%. Condenser(G1) peaks at 183.6 mbar. MC(H35) → 0.27 K. Still(P1) → 1.12 K / 0.6 mbar.
1510 V6 → 100%. Condenser(G1) peaks at 179.4 mbar. MC(H35) → 0.137 K. Still(P1) → 1.06 K / 0.17 mbar.
1521 Plug in DCRCs 1 and 3
1522 MC = 50mK, G1 = 154mbar
1554 Open V14 to pull the rest of the mixture out of the dump into circulation.
MC rises to 66 mK, Still peaks above 30 mbar then falls again, Condenser rises to 148 mbar. 1K Pot unaffected.
as usual we'll wait till Still is below 1 mbar then close V14.
1541 Begin LHe transfer @ MB = 33.5%
1610 Fill LN2 cold trap dewar
1617 Still = 0.8 mbar. close V14
1624 End LHe transfer @ MB = 92.0%
1710 Start DAQ programs and check phonon channels, all working. Most baselines are still near -0.8 V, except DCRC3pD.
1720 Adjust phonon offsets to get all channels near -0.80 V
1725 Dial HV up to -109.25V nominal, -107.1V actual, for 15% prebias (using RootSequencer)
1731 Dial HV down to -95V nominal, -92.8V actual
1735 HV = -92.8 V, Vtest = -22.3 mV, Ileak = -24.5 uA, Vdet = -83 V (89%). Leakage is a little high
1735 Dial HV down to -94V nominal, -91.9V actual
1735 HV = -91.8 V, Vtest = -20.9 mV, Ileak = -23 uA, Vdet = -82.6 V (90%).
1735 MC = 29.4 mK
1741 Start -82.6 V PuBe data series 46 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
NOTE: the NaI trigger window has been changed. We had been using an upper threshold of 11.0 MeV thinking that was low enough to be below saturation and therefore cut out muons. But we took a NaI spectrum scan recently and found that some detectors have thresholds below 11.2 MeV. So we've reduced the upper edge to 11.0 MeV. Should reduce trigger rate from around 11 Hz to around 8 Hz.
1742 HV = -91.7 V, Vtest = -20.8 mV, Ileak = -22.9 uA, Vdet = -82.6 V (90%).
1832 Start -82.6 V PuBe data series 202 min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
2157 Start -82.4 V PuBe data series 42 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2200 HV = -91.7 V, Vtest = -21 mV, Ileak = -23.1 uA, Vdet = -82.4 V (90%).
2240 Start -82.5 V PuBe data series 533 good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
DAQ crash at 7:42 AM, dumps after 533 are empty
2240 HV = -91.7 V, Vtest = -20.8 mV, Ileak = -22.9 uA, Vdet = -82.5 V (90%).

Next Day