start › cdms › k100 › run_diary › 20220923

2022/09/23 (Friday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Warmed up yesterday for ninth 5K pumpout. Turbo pumping IVC overnight.
1205 MC/1KPot/Sorb = 5.40/6.57/8.30 K. G1(fridge) = 545 mbar
1205 Turbo on IVC = 1.2 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
1210 Close IVC valves on top of fridge
1210 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.25 bar on mechanical gauge.
1210 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1215 Start pumping 1K Pot continuously
1215 MC/1KPot/Sorb = 5.40/3.37/4.2 K.
1235 MC falls quickly…
1235 MC/1KPot/Sorb = 3.44/3.22/4.0 K. G1(fridge) = 470 mbar
1310 MC/1KPot/Sorb = 3.24/2.88/4.0 K. G1(fridge) = 374 mbar
1310 1K Pot = 0° / 2.88 K / 2.6 mbar
1310 Open 1K Pot needle valve → 15°. 1K Pot → 2.57 K / 6.0 mbar
1310 MC = 3.15 K. G1(fridge) = 335 mbar
1315 Open 1K Pot needle valve → 30°. 1K Pot → 2.33 K / 8.0 mbar
1325 MC = 2.99 K. G1(fridge) = 278 mbar
1325 Open 1K Pot needle valve → 45°. 1K Pot → 2.10 K / 13 mbar
1350 Fill LN2 cold trap bath
1400 Some kinks in the MC cooling curve around 2.1 K, around the helium lambda point.
1410 Remove PuBe source from PAN43 K100 lab and put in storage in PAN443. We'll take data without the source this weekend to better understand the calibration and the NaI-coincident spectrum.
1415 MC = 2.08 K. G1(fridge) = 55 mbar. These are both flattening.
1420 Sudden drop in MC from 2.08 to 1.99 K, maybe that's the lambda point.
1420 1K Pot = 1.54 K / 7.0 mbar, kinda low. Last time though it went up at some point, we'll try starting circulation.
1420 Close V3 and V9, leaving 49 mbar mixture in dump (G2)
1420 Turn on 3He pump. G1(condenser) → 57 mbar.
1430 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 310 mbar. MC(H35) → 1.41 K. Still → 1.44 K / 26 mbar.
1K Pot → 1.60 K / 14 mbar, better.
1435 V6 → 13.0%. Condenser(G1) peaks at 236 mbar. MC(H35) → 1.15 K. Still(P1) → 1.19 K / 5.5 mbar.
MC wiggles around 1.2 K, Al Tc.
1450 V6 → 13.4%. Condenser(G1) peaks at 194 mbar. MC(H35) → 1.08 K. Still(P1) → 1.18 K / 3.6 mbar.
1500 V6 → 13.8%. Condenser(G1) peaks at 176 mbar. MC(H35) → 0.72 K. Still(P1) → 1.16 K / 2.4 mbar.
1505 V6 → 14.4%. Condenser(G1) peaks at 179 mbar. MC(H35) → 0.42 K. Still(P1) → 1.14 K / 1.10 mbar.
1515 V6 → 15.0%. Condenser(G1) peaks at 177 mbar. MC(H35) → 0.20 K. Still(P1) → 1.15 K / 0.45 mbar.
1525 V6 → 100%. Condenser(G1) peaks at 161 mbar. MC(H35) → 56 mK. Still(P1) → 1.11 K / 0.10 mbar.
1550 MC = 33.6 mK. Condenser(G1) = 121 mbar. Still = 0.075 mbar.
1555 Accidentally open V4 instead of V14. That let some mix from the Condenser into the Dump. Close V4.
1555 Open V14 to pull the rest of the mixture out of the dump into circulation. MC peaks at 98 mK, Still peaks at 260 mbar, Condenser peaks at 203 mbar, then all relax again. 1K Pot unaffected.
1615 Still = 0.85 mbar.
Close V14
1625 MC = 31.7 mK and cooling
1635 MC = 28.9 mK
1635 Plug in DCRCs 1&3
Starting the DAQ programs didn't work as usual - the DAQ programs started without shutting themselves off even with all DCRCs enabled, that's not the usual behavior. But I couldn't start a series without a crash. Wound up restarting MIDAS, now it's working ok.
1700 Check phonon channels. They look good. Baselines still near -0.8 V except DCRC3pD, had to adjust its offset from 1.37 to 1.65
1700 NaI is still triggering without the PuBe, rate is not severely reduced. We'll see when we take data.
1705 Fine tune offsets to get baselines close to -0.80 V
1705 MC = 29.4 mK. Seems to have settled here after turning on the DCRCs.
1709 Start 0 V bg data series ## min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
1746 Start 0 V bg data series ## min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button