start › cdms › k100 › run_diary › 20220903

2022/09/03 (Saturday)

Previous Day

  • Lab in-personnel:Giulia, Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
fourth 5K pumpout yesterday. Pumping IVC overnight.
1440 MC/Coldplate = 9.53/19.6 K, Still = 14.7 K, G1/G2 = 558/502 mbar, 1K Pot = 6.99 K, MB = 68.2 % .
1450 Fill LN2 cold trap bath.
1450 Turbo on IVC = 1.4 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
1450 Close IVC valves at top of fridge.
1455 Turn on 4He pump. Open Recovery valve to 4He pump. Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.19 bar on mechanical gauge.
1455 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low.
1500 Start pumping 1K Pot continuously
1505 MC = 9.51 K, cooling slowly. 1K Pot = 3.64 K, cooling. Sorb = 4.3 K, cooling.
1510 1K Pot = 3.40 K, cooling
1515 Open 1K Pot needle valve → 15°. 1K Pot → 3.17 K / 5.0 mbar
1520 Open 1K Pot needle valve → 30°. 1K Pot → 2.95 K / 9 mbar
2000 MC = 2.4 K, G1(Fridge) = 110 mbar, 1K Pot = 1.57 K / 10 mbar
2000 Turn on 3He pump
2010 Close V3 and V9, leaving 110 mbar mixture in dump
2010 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 520 mbar. MC(H35) → 1.60 K.
2015 V6 → 13.0%. Condenser(G1) peaks at 433 mbar. MC(H35) → 1.42 K. Still(P1) → 21 mbar.
2025 V6 → 13.4%. Condenser(G1) peaks at 308 mbar. MC(H35) → 1.31 K. Still(P1) → 7 mbar.
2030 V6 → 13.8%. Condenser(G1) peaks at 316 mbar. MC(H35) → 1.23 K. Still(P1) → 4 mbar.
2035 V6 → 14.4%. Condenser(G1) peaks at 310 mbar. MC(H35) → 1.15 K. Still(P1) → 2.5 mbar.
2040 V6 → 15.0%. Condenser(G1) peaks at 322 mbar. MC(H35) → 0.80 K. Still(P1) → 2 mbar.
2050 V6 → 100%. Condenser(G1) peaks at 328 mbar. MC(H35) → 0.25 K. Still(P1) → 0.3 mbar.
2055 Open V14 to pull the remaining mixture out of the Dump and into circulation
Condenser peaks at 510 mbar then falls again. Still goes off-scale then falls again. MC rises from 0.13 K to 0.2 K then falls again.
2105 1K Pot fails, → 2.15 K, pressure falls from 17→10 mbar. Must have run dry. Probably overwhelmed it trying to condense all that mix at higher than usual temperature. Plus we only have the needle valve open to 30°
G1 = 480 mbar and rising. MC reached minimum 80 mK, now rising.
2120 Still = 1.5 mbar. Close V14.
2120 Condenser = 745 mbar and rising, 1KPot = 2.7 K, MC = 95 mK. Close V6 to stop circulation.
2130 Condenser = 330 mbar and falling. 1KPot = 2.15 K and falling. MC = 1.0 K.
2200 MC = 1.47 K. 1K Pot = 1.53 K / 6.5 mbar. Condenser(G1) = 72 mbar.
Note: forgot to turn off DCRCs before cooling to base.