  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
Fridge failure six days ago, looked like an impedance blockage. At 110 K the impedance blockage was gone. Recooling: did initial LHe transfer yesterday
1125 MC(Cernox)/Sorb = 4.8/9.0, G1(fridge) = 551 mbar. MB = 28.9% / 126 mbar+1atm
1125 Turbo on IVC = 2.2 nbar. With Turbo valved out for 1 minute IVC = 19 nbar
1139 Begin LHe transfer @ MB = 28.9%
1145 Fill LN2 cold trap bath
1150 Close V1a, then pump out line leading to 1K Pot with scroll pump. Then switch over to 4He pump. Turn on 4He pump then change pump exhaust to recovery and open valve to recovery lines.
1155 Check pressure in 1K Pot: Mechanical gauge = 0.43 bar. (Note: V4a was closed during this check.) 1K Pot temperature dips briefly to 3.3 K
1200 Pump out 1K Pot in steps, alternating between V4a and Speedivalve. Throttle valves to keep pressure at P2 below around 10 mbar. After 6 iterations, leave both valves open.
1205 MC(Cernox)/MC(H35)/Goldman = 4.71/4.71/4.64 K. (Coldplate/Still = 2.800/3.006 kΩ = 7.36/3.33 K. These don't have good calibrations here.)
1210 Sorb = 4.0 K. G1 (fridge) = 399 mbar and falling. 1K Pot = 2.96 K / 2.0 mbar
1215 Open 1K Pot needle valve → 15°. 1K Pot → 2.5 K / 11 mbar.
1220 Open 1K Pot needle valve → 30°. 1K Pot → 2.1 K / 24 mbar. MC(Cernox) plummets to 3.3 K.
1228 End LHe transfer @ MB = 93.4%
1235 MC(Cernox) = 2.59 K and falling. 1K Pot = 1.62 K / 20 mbar
1235 Open 1K Pot needle valve → 40°. 1K Pot → 1.65 K / 32 mbar.
1245 1K Pot = 1.57 K / 13 mbar, both falling
1250 MC (H35) = 2.08 K
1250 Open 1K Pot needle valve → 50°. 1K Pot → 1.58 K / 14 mbar.
1255 MC (H35) = 2.02 K, cooling slowly. G1(fridge) = 57 mbar, flattening
1300 1K Pot pressure jumps - lambda point probably. Jumps off scale then waves up and down between 60 and 180 mbar
2100 MC (H35) = 1.83 K, G1(fridge) = 45 mbar. 1K Pot = 50° / 1.71 K / 80±10 mbar.
2105 Close V3 and V9, leaving 50 mbar (G2) in the dump. Turn on 3He pump.
2105 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 170 mbar. MC(H35) → 1.2 K, then bounces back up to 1.35 K and rising. 1K Pot pressure spikes off-scale then recovers. P1 (still) → 28 mbar.
2115 V6 → 13.0%. Condenser(G1) peaks at 143 mbar. MC(H35) wiggling, went up to 1.4, now 1.25 K. Still(P1) → 9.0 mbar, although it spiked at the same time MC peaked.
2130 V6 → 13.4%. Condenser(G1) peaks at 140 mbar. MC(H35) → 1.1 K. Still(P1) → 4.5 mbar.
2135 V6 → 13.8%. Condenser(G1) peaks at 135 mbar. MC(H35) → 0.58 K. Still(P1) → 3.2 mbar.
2150 V6 → 14.5%. Condenser(G1) peaks at 135 mbar. MC(H35) → 0.40 K. Still(P1) → 1.4 mbar.
2200 V6 → 15.0%. Condenser(G1) peaks at 122 mbar. MC(H35) → 0.19 K. Still(P1) → 0.6 mbar.
2220 V6 → 100%. Condenser(G1) peaks at 110 mbar. MC(H35) → 42 mK. Still(P1) → 0.14 mbar.