2022/07/13 (Wednesday)
- Lab in-personnel: Matt, Giulia
| Status | Run 76 in progress |
|---|---|
| Setup | Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4) Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors. PuBe source behind lead wall. NaI trigger window -t10760 -T15760 (7.6-11.2 MeV) |
| Taking 0V NaI-triggered data continuously | |
| 1040 | MC/Coldplate = 29.1/155 mK, Still = 1.12 K / 0.095 mbar / 0 mW, G1/G2 = 106/102 mbar, 1K Pot = 60° / 1.55 K / 8.0±0.5 mbar, MB = 42.5% |
| 1051 | Filled LN2 cold trap bath |
| 1505 | MIDAS web interface crashed at some point. Restart MIDAS. |
| 1603 | Begin LHe transfer @ MB = 35.2% |
| 1652 | End LHe transfer @ MB = 95.7% |