2022/07/07 (Thursday)

Previous Day

  • Lab in-personnel: Giulia, Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall.
NaI trigger window -t10760 -T15760 (7.6-11.2 MeV)
0632 Start -197.3 V PuBe data series 7 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
0635 MC = 29.2 mK
0639 Start -197.3 V PuBe data series 3 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
0640 Crap - discovered that DCRC4 is not enabled. I guess maybe it became disabled when Tower_FE crashed yesterday??
Hmm, according to ODB records, all series started yesterday had DCRC4 enabled. Good, then.
0644 Start -197.3 V PuBe data series 3 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
This time DCRC4 is enabled
1105 MC/Coldplate = 29.1/152 mK, Still = 1.08 K / 0.095 mbar / 0 mW, G1/G2 = 133/128 mbar, 1K Pot = 60° / 1.55 K / 8.0±0.5 mbar, MB = 37.6%.
1115 Didn't take the Vtest before attempting(flow was also too low) the LN2 transfer, and right after that it was oscillating betwen -20mV and -40mV.
1135 Filled LN2 cold trap dewar.
1145 Vtest = -10 mV
1147 Dial HV down to 0V
1539 Begin LHe transfer @ MB = 31.4%
1630 End LHe transfer @ MB = 87.1%
1643 Start 0 V PuBe data series 27 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
Put extra tap around Ebox purge bag to try to close off gaps
1710 Adjust LN2 Dewar port valve to get just a little flow of N2 gas for purge
1720 Dial HV from 0V up to -230V
1725 Dial HV from -230V down to -200V nominal. True HV = -197.3 V
1727 Vtest ~ -21 mV
1730 Found more big gaps in Ebox purge bag. Tape those up.
1732 Vtest ~ -30 mV
1740 Rig up wires so we can read Vtest with the Keithley meter, and monitor it with LabVIEW.
1740 Vtest = -33.3 mV. Ileak = -37 nA, Vdetector = -182.6 V (93%)
1748 Start -197.3 V PuBe data series 9 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
1758 Start -197.3 V PuBe data series 229 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
1800 MC = 28.7 mK
2149 Start -197.3 V PuBe data series 14 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
2205 Start -197.3 V PuBe data series 611 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
Vtest was between ~-30 and -32 mV throughout series