UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20220331

2022/03/31 (Thursday)

Previous Day

  • Lab in-personnel: Anthony, Kitty, Matt
Status Run 76 in progress
Setup Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4)
1450 MC/Coldplate = 28.6/140 mK, Still = 1.12 K / 0.120 mbar / 0 mW, G1/G2= 176/177 mbar, 1K Pot = 18° / 1.64 K / 24±2 mbar. MB = 28.9%
1450 Goldman = 72.7 mK @ 632 uB excitation
1606 Begin LHe transfer @ MB = 27.0%
1644 End LHe transfer @ MB = 95.3%
1705 Fill LN2 cold trap bath
1715 Solder a LEMO cable to a Dsub50 diverter board, DCRC-side (male connector), with center conductor on Pin 42 and shield soldered to Dsub shell/ground. Pin 42 is LED1. The intent is to use this to feed a TTL signal into a DCRC, and set the DCRC to connect LED output to pA - then we can trigger on the external TTL signal. Something like this was done back in Run 62.