| Status | Run 76 in progress |
| Setup | Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4) |
| 0840 | MC/Coldplate = 28.8/140 mK, Still = 1.12 K / 0.130 mbar / 0 mW, G1/G2= 178/166 mbar, 1K Pot = 18° / 1.66 K / 27±2 mbar |
| 0935 | Do mini-IbIs on DCRC1pA/pB to find QET bias for dP/dI = 0.7 uV |
| 1110 | Do mini-IbIs on DCRC1pC/pD to find QET bias for dP/dI = 0.7 uV |
| 1140 | Do mini-IbIs on DCRC3pA/pB to find QET bias for dP/dI = 0.7 uV |
| 1350 | Attempting the same on DCRC3pC. But the lowest dP/dI before it falls into superconducting is 0.95 uV. And there are a lot of electrothermal oscillations near the bottom of the transition. Set to dP/dI = 1.02 uV. |
| 1405 | Do mini-IbIs on DCRC3pD to find QET bias for dP/dI = 0.7 uV |
| 1435 | Do mini-IbIs on DCRC4pA/B to find QET bias for dP/dI = 0.7 uV |
| 1510 | Do mini-IbIs on DCRC4pC/D to find QET bias for dP/dI = 0.7 uV |
| 1510 | MC = 28.7 |
| 1525 | Redo mini-IbIs on DCRC1pB to find QET bias for dP/dI = 0.7 uV, and extend so that we can also set to 1.0 uV |
| 1536 | Begin LHe transfer @ MB = 19.8% |
| 1624 | End LHe transfer @ MB = 93.1% |
| 1630 | Fill LN2 cold trap bath |