start › cdms › k100 › run_diary › 20220218

2022/02/18 (Friday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 finished. Began warmup two days ago. 1K Pot pumped dry and isolated. All mixture in dump, fridge isolated. LHe Trap and LN2 Trap 1 still have contaminants. Some He gas in IVC and in Cryostat Vacuum Jacket. Main bath empty and disconnected from recovery.
1330 MC(Cernox) = 252.1 K, warming at 3.2 K/hr. Taking T_room to be 295.3 K this implies warming tau of 13.5 hours. Taking dew point to be 273 K this implies we'll reach dew point in 8.9 hours, which is tonight at 10:25 pm.
1340 Main bath (G3) = 323 mbar+1atm. That's not that high, so probably can just leave it sealed. It was higher earlier, not sure why it's falling but whatever.
1415 MC(Cernox) = 254.6 K. Slowly let N2 into IVC ending at about 0.87 bar, which should be close to 1 atm at room temperature.