start › cdms › k100 › run_diary › 20220116

2022/01/16 (Sunday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 30.03 mm
1730 Take some mini-IbIs at Goldman = 44.8 mK. Then set QET biases to target dP/dI=1.0 uV and adjust driver offsets so baselines ~ -0.9 V
1746 Start 0V data series, 25 min long, with 1000 initial randoms, no in-run randoms, trigger on DCRC3pB threshold 10.
Start series with the Start Button.
1750 Goldman = 44.7 mK
1755 Goldman/Tower/MC(H35)/Coldplate = 44.7/121/33.5/160 mK, Still = 1.13 K / 0.110 mbar / 0 mW, G1/G2= 254/242 mbar, 1K Pot = 1.61 K / 15±1 mbar, MB = 20.3%.
1813 Begin LHe transfer @ MB = 19.8%
1820 Fill LN2 cold trap bath
1852 End LHe transfer @ MB = 90.4%

Next Day