2021/12/23 (Thursday)

Previous Day

  • Lab in-personnel: Matt, Himangshu
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place, S wall partially complete
1115 Goldman/Tower/MC(H35)/Coldplate = 46.5/89.2/35.2/209 mK, Still = 1.71±0.04 K / 0.130 mbar / 0 mW, G1/G2= 199/198 mbar, 1K Pot = 1.63 K / 21±1 mbar, MB = 30.9%.
1125 Fill LN2 cold trap bath.
1135 Perform mini-IbIs on DCRC1pA. The QET bias had been set to 241 uA to target a dP/dI of 1.0 uV, based on IbIs data from a couple weeks ago when the fridge was warmer. IbIs now suggests a dP/dI of 0.77 uV.
1156 Start 0V data series, 29 min long, with 1000 initial randoms, no in-run randoms, trigger on DCRC3pB threshold 5.
Start series with the Start Button.
We'll set up to measure some Tc's. First, let's turn all the QET biases to zero. We'll wait a bit to see if this causes a noticeable change in tower temperature.
1320 Set all QET biases to 0. Before this they were at: DCRC1/3/4 = 241,319,275,184 / 163,165,216,0 / 201,145,265,186
Side 2 channels seem to have lower Tc's so let's start with DCRC4 which has three Side2 channels.
1340 Set up DCRC4 for Tc tests. Set Test Signal to 12 mVpp / 200 Hz. Trigger on test signal.
Connecting test signal to QET seems to work on pB, pC, pD but not pA. Note that this is done with the SQUID tuning GUI - I don't know how to do it on the ODB.
1410 Watching DCRC4pB and pD mostly (pC seems to have higher Tc according to IbIs), start applying heat to MC. Start with 1 uW. Goldman = 46.4 mK before applying heat.
1422 MC = 5 uW, Goldman = 64.5 mK. No transitions seen yet
1431 MC = 22 uW, Goldman = 73.3 mK. pD transitioned was 66.5 mK. pB transition was 70.4-70.5 mK.
1440 MC = 30 uW, Goldman = 89 mK. pC transition was 78.1-81.9 mK. Start cooling now.
1455 MC = 0 uW, Goldman = 67 mK. All channels are back in s/c state.
DCRC4pB (Side2pA) DCRC4pC (Side1pA) DCRC4pD (Side2pD)
Transition start, warming 70.4 78.1 66.5
Transition end, warming 70.5 81.9 66.5
Transition start, cooling 74.8 85.9 73.6
Transition end, cooling 74.3 81.5 ~73
1455 Fill LN2 cold trap bath.
1504 Begin LHe transfer @ MB = 24.9%
1520 Restore transition QET biases on all 11 channels and adjust driver offsets to set baselines ~ -1.0 V
1538 End LHe transfer @ MB = 83.2%

Next Day