start › cdms › k100 › run_diary › 20210927

2021/09/27 (Monday)

Previous Day

  • Lab in-personnel: Giulia and Matt
Status Run 74 finished. Warming up.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. Lead shield erected, 4 walls. Currently both DCRCs are on.
All mixture is in dump. Fridge open to IGH. V4a closed so no longer pumping on 1K Pot.
1006 Tower/MC(H35)/MC(Cernox)/Sorb = 10017/NaN/10634/30900 mK
1017 We closed the manual V12b valve.
1022 We closed the 1KPot speedivalve.
1325 Close valve at 4He pump inlet and turn off 4He pump.
1330 Pump out 1K Pot with scroll pump, then close 1K Pot speedivalve
1335 Turn off DAQ programs (triggerfe, towerfe, dcrc_driver) and power down DCRCs.
1340 Pump out fridge with Scroll Pump (? I don't think we did this. -Matt)
1340 Tower/Sorb = 11.9/39.0 K
1350 Let about 0.4 mbar He gas into IVC. Tower temperature rises above 16 K. Sorb rises above 43 K.
1410 Tower = 18.8 K, flat. Sorb = 42.9 K and falling.
1445 Isolate IVC, then move pump manifold line over to cryostat vacuum jacket(CVJ) port
1450 Let 4.6 mbar He gas into vacuum manifold, then let this into CVJ. CVJ → 74 μbar
He recovery rate increases from 12 to 40 cuft/hour.
1500 Remove lead/poly shield