2021/09/16 (Thursday)
- Lab in-personnel: Giulia and Matt
| Status | Run 74 in progress. First cooldown after IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/ |
|
| Position of Am source: y = 0.0 mm | |
| Circulation is suspended so we can explore data at higher temperatures. 1K Pot is still on. | |
| 0838 | Tower/MC/Coldplate = 1682/11629/3057 mK, Still = 2.27 K / 1000.0 mbar / 0 mW, G1/G2= 32800/33600 mbar, 1K Pot = 1.56 K / 10±1 mbar. MB = 75.2%. |
| 0844 | Fill LN2 cold trap bath |
| 0900 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 3.2/1.5/2.0/29 8e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 550/3.7/4.1/14 e-6. |
| Noise is lower than yesterday. | |
| 0930 | Set MC heater → 2000 uW |
| 1030 | Tower temperature hasn't changed much Increase MC heater → 5000 uW |
| 1100 | Tower = 1.92 K, seems to be leveling |
| 1100 | Increase MC heater → 10000 uW |
| 1200 | Tower = 2.10 K, flat |
| 1202 | Start 45 min series of +10/0V data (bias applied after 60 sec), at elevated temperature, with 1000 beginning randoms, trigger threshold 50/25/30/120 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series. |
| 1520 | Increase MC heater → 20000 uW |
|---|---|
| 1530 | Tower = 2.7 K and rising in a sort of runaway fashion… Turn off MC heater for now |
| 1640 | Have been messing around with different heater settings, seems a little difficult to keep it stable in this temperature regime (2.1-2.8 K) with the 1K Pot on at least. |