2021/08/13 (Friday)

Previous Day

  • Lab in-personnel: Giulia and Matt
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
Position of Am source: y = 0.00 mm
0908 Tower/MC/Coldplate = 109.9/35.5/ 190.6 mK, Still = 1.35 K / 0.075 mbar / 0 mW, G1/G2= 280.1/253.4 mbar, 1K Pot = 1.59 K / 20±1 mbar. MB = 72.4%.
0921 Fill LN2 cold bath.
1115 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 24/4.0/2.3/43 8e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 660/69/60/1500e-6
Noise at 10kHz is a little lower today but 3kHz oscillation is higher today than yesterday
1125 Place Na-22 source in lead castle at detector level 80 cm from detector center (54.5 cm from mu-metal).
1129 Start 50 min series of 0V/-12V (bias applied after 60 sec), with Na-22 source at 80 cm, with 1000 beginning randoms, trigger threshold 120/40/60/200 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1222 Dial HV down to 0V and flash detector for 10 minutes with DCRC1&3 LEDs.
07210813_1129
1408 Start 48 min series of 0V/-9V - 0V/-12V (bias applied after 60 sec, then 3 minute wait, then increase to -12), with 1000 beginning randoms, trigger threshold 100/50/60/200 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1500 Dial HV down to 0V and flash detector for 10 minutes with DCRC1&3 LEDs.
07210813_1408
1603 Start 45 min series of 0V/-12V - 0V/0V (bias applied after 60 sec, then 15 minute wait, then decrease to 0V), with 1000 beginning randoms, trigger threshold 100/50/60/200 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1648 Flash detector for 10 minutes with DCRC1&3 LEDs.
07210813_1408