Previous Day

  • Lab in-personnel: Giulia and Matt
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
Position of Am source: y = 0.00 mm
0006 Move sequence ends
0230 Tower cools to within 10 mK of base
0300 Tower cools to within 5 mK of base
0400 Tower cools to within 1 mK of base
0908 Tower/MC/Coldplate = 79.5/35/ 190.2 mK, Still = 1.37 K / 0.08 mbar / 0 mW, G1/G2= 270.4/255.7 mbar, 1K Pot = 1.59 K / 20±1 mbar. MB = 28.6%.
0919 Fill LN2 cold bath.
0921 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 2.2/2.0/2.2/6.4 8e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 9.2/1.4/3.3/18 e-6
Noise is slightly higher than yesterday.
1030 Trying to see whether we can monitor test resistor while taking data. In general, looks like connecting a BNC to the test resistor jack increases noise by an order of magnitude, and connecting that to the Keithley meter increases it another order of magnitude.
1035 Try HP meter instead - this is worse, gives some weird oscillating pulses.
1045 Plugging into Keithley adds noise only if the Keithley is plugged in. Replugging it into a “clean” outlet or trying to insulate its ground plug doesn't help.
1050 Set triggers, need to set them above 1000.
The plan here is to do another -15V/0V series. We've seen that there's a dropout in the impact ionization in the peak when we set the bias this high. We haven't seen any corresponding change in bias voltage. So we want to see if there's any change in leakage current. So - we'll take data with the Keithley reading leakage and the HP reading bias. But at these high thresholds I don't think we'll see the 60keV line.
1100 Start 46 min series of 0V/-15V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 4000/4000/4000/4000 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
I started this series with trigger thresholds 1500-3000 but I didn't see any real looking pulses so I increased to 4000. Then I saw some pulses.
1105 Vtest was 72 μV before bias was applied. When bias was ramped up, Vtest dropped briefly to 62 μV then settled to a 68 μV baseline. Then 3 minutes after the bias was applied, there was a sudden change in Vtest, fell to -6 μV. Then it relaxed exponentially back to 68 μV with a tau of 51 seconds.
if we assume some offset, and 1 kΩ test resistor, then leakage is around 4 nA baseline, and the deviation peaked at 74 nA.
1150 Dial HV to 0V. Vtest goes back to 71-72 μV.
1150 Flash detector for 10 minutes with DCRC1&3 LEDs.
1539 Start LHe transfer @ MB = 17.4%
1633 End LHe transfer @ MB = 91.4%

20210810_log.xlsx