  • Lab in-personnel: Matt
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
1105 Tower/MC/Coldplate = 86.7/35.0/171 mK, Still = 1.10 K / 0.075 mbar / 0 mW, G1/G2= 302/275 mbar, 1K Pot = 1.63 K / 20±1 mbar. MB = 66.5%
1110 Fill LN2 cold trap bath.
1110 Do mv -32 with dwell = 40 and power = 6.8 V. This should move source back to the center, y = 0.00 mm.
1120 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 2.0/1.2/1.4/4.9 e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 19/3.2/2.4/32 e-6
This is much lower noise than yesterday, the 3kHz oscillations are down by a factor of 50. Like last week.
Set things up to measure event rate. We'll increase the trace length then we'll apply bias and count pulses.
1135 Increase ChargeWaveFormLength in DCRC1&3 from 8192→65536. This will increase the trace length from 1.6384 ms to 13.1072 ms.
1140 Apply -15V bias to Side 2. Now we are seeing some pulses.
1145 Event rate in DCRC1qI(q4) for pulses above 0.1V: 12 pulses counted in 60 traces. 15.2 Hz.
1150 Event rate in DCRC3qO(q1) for pulses above 0.1V: 4 pulses counted in 60 traces. 5.1 Hz.
1155 Assemble a stand and a lead castle for external sources. Source will be at the same vertical level as the detector (no - actually about 8cm lower than detector level), with opening in the castle toward the detector (source is N of detector). Mark spots on floor at 20cm intervals for distance placement.
1215 Dial bias down to 0V
Flash the detector for 60 sec.
1340 Apply -15V bias to Side 2. Seeing pulses again.
1350 Place Ba-133 source 1.40 m, 1.0 m, 0.6 m from the center of the detector. Count pulses in 60 traces in each case: 11,5,9. So no increase in rate, this is not close enough.
Waiting for a change at IT, computers down. Dial bias to 0V.
1530 Flash detector for 60 seconds.
1535 Place Ba-133 source 0.40 m from center of detector
1540 Adjust height of source stand, now source is at nearly same level as detector.
1545 Apply -15V bias to Side 2.
1550 Event rate in DCRC1qI(q4) for pulses above 0.1V: 34 pulses counted in 120 traces. 21.6 Hz.
1555 Event rate in DCRC3qO(q1) for pulses above 0.1V: 45 pulses counted in 120 traces. 28.6 Hz.
1600 Put Ba-133 source back in safe
1600 Place Na-22 source 0.40 m from center of detector
1605 Event rate in DCRC3qO(q1) for pulses above 0.1V: 53 pulses counted in 120 traces. 33.7 Hz.
1610 Event rate in DCRC1qI(q4) for pulses above 0.1V: 31 pulses counted in 120 traces. 19.7 Hz.
1610 Dial bias down to 0V and flash detector for 60 seconds
1610 Restore ChargeWaveFormLength to standard settings.
1646 Start series but having trouble with triggering, abort.
1700 Start series but having trouble with triggering, abort.
1711 Start series but having trouble with triggering, abort.
1718 Start 45 min series of 0V/-15V (bias applied after 60 sec), with Na-22 source at 0.40 m, with 1000 beginning randoms, trigger threshold 120/90/40/70 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
oops - DCRC1 triggers were left off for this series.
1807 Dial bias down to 0V and flash detector for 60 seconds.