UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20210626

2021/06/26 (Saturday)

Previous Day

  • Lab in-personnel: Derek
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. (actually currently DCRC55(1) is disconnected. DIB2+acc→SL43→filter→cable3→Source_mover_board)
1047 Tower/MC/Coldplate = 79.0/30.3/170 mK, Still = 1.07 K / 0.080 mbar / 0 mW, G1/G2= 209/201 mbar, 1K Pot = 1.64 K / 24±2 mbar. MB= 57.7
Coldplate and Still thermometers are in their “low mode” today, they we in “high mode” yesterday.
1048 Fill LN2 cold trap bath.

Next Day