| Status | Run 74 in progress. First cooldown after IVC rebuild. Started circulation last week. |
| | Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. |
| | Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. |
| 1400 | Tower/MC/Coldplate = 146/32.4/191 mK, Still = 1.38 K / 0.085 mbar / 0 mW, G1/G2= 167/162 mbar, 1K Pot = 1.64 K / 24±2 mbar. |
| | This is a lower Tower reading than we've seen before. Coldplate and Still are reading high, we've seen this before - both of these thermometers (which share a fridge cable) will show bimodal behavior. |
| 1424 | Begin LHe transfer @ MB = 29.5% |
| 1435 | Begin LN2 transfer. |
| 1450 | Ended LN2 transfer. |
| 1500 | Ended LHe transfer @ MB = 90.1% |
| 1515 | Zach and Gabe leave the lab. |