2021/05/07 (Friday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Derek
Status Run 74 in progress. Warming up after 3rd appearance of IVC leak at 4K. Helium in Cryostat Vacuum Jacket, mixture in fridge
1000 MC(Cernox)/Sorb/1K Pot = 95.0/192/>7 K, G1 (fridge) = 607 mbar
1000 IVC = 190 mbar.
1000 Start pumping mixture from fridge and LN2 Trap 1 to Dump
1000 MC = 95.3 K, rising 4.8 K/hr, suggests warming tau=41.6 hrs, projected to reach dew point Tuesday May 11 at 8 am. Probably should put more gas in CVJ.
1200 Matt enters lab
1200 IVC = 190 mbar, 0.31 bar on mechanical gauge
1200 MC(Cernox) = 104.1 K
1205 Stop pumping mixture to dump. Dump (G2) = 730.8 mbar. (Later at 12:40, before turning off the 3He pump, it had risen to 731.8 mbar.)
1205 Again compare gas from IVC to N2 by comparing the readings on the two gauges - it looks like N2, must be very little He (because pressures above 0.05 bar or so of He cause the Pfeiffer gauge to saturate)
1220 Start gently pumping IVC with Scroll Pump. This stalls the warming of the MC.
1230 IVC < 0.1 mbar, switch to pumping with Turbo
1235 Recovery = 0.18 cuft/hr. At a guess, Recovery = 433239 cuft.
1305 Turbo on IVC = 19 nbar. With Turbo valved out for 1 minute, IVC = 0.25 μbar.
1315 Slowly fill IVC with 0.36 bar N2
1335 MC(Cernox)/Sorb = 109.0/224 K
1340 CVJ = 0.054 mbar
1340 Let 8.6 mbar N2 gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket. Vacuum Jacket → 0.30 mbar.
1405 Recovery = 0.32 cuft/hr.
1445 MC = 115 K, rising 5.7 K/hr, suggests warming tau=31.4 hrs, projected to reach dew point Monday May 10 at 10:26 am.
1505 Close recovery valve from MB outlet
1515 Matt exits lab
1600 MC = 123 K, rising 6.4 K/hr, suggests warming tau=26.9 hrs, projected to reach dew point Monday May 10 at 00:39 am.