UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20210502

2021/05/02 (Sunday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Monica
Status Run 74 in progress. Warming up after second occurrence of IVC leak during initial LHe transfer. Helium in Cryostat Vacuum Jacket. N2 in IVC. Mixture in dump.
1545 Matt enters lab
1545 MC(Cernox) = 276 K.
1545 IVC = 0.75 bar on mechanical gauge. Let's put more N2 into it.
1550 Slowly fill IVC with N2 → 0.93 bar on mechanical gauge
1555 Close off IVC then switch pumping line over to CVJ and pump out line
1555 Check pressure in CVJ: 0.19 mbar
1555 Begin pumping out CVJ, with Scroll for a minute than @ <0.1 mbar switch to Turbo
1600 Turbo on CVJ = 1.2 μbar.
1600 Matt exits lab