2021/04/20 (Tuesday)
- Lab in-person: Matt
- Virtual partner: Zach
| Status | Run 74 in progress. Warming to fix IVC leak. Helium gas in cryostat vacuum jacket. Mixture in dump. Recovery valve open. |
|---|---|
| 0705 | MC(Cernox)/Sorb = 118/250 K |
| 1105 | MC(Cernox) = 141 K, rising at 5.4 K/hour. Suggests tau_warming of 28.6 hours, predicted to reach dew point Thursday night at 19:50. |
| 1105 | IVC = 300 mbar on digital gauge. Could be overpressurized now, need to go in and check. |
| 1335 | Matt enters lab |
| 1335 | MC(Cernox) = 153.9 K |
| 1335 | IVC = 300 mbar on digital gauge. Check pressure on mechanical gauge: a little less than 1 atm, maybe 0.95 bar. Good, still not overpressured. |
| 1335 | Pump out IVC gently with Scroll pump for 5 minutes, down to about 0.3 bar. Notice that this flattens out the temperature, apparently a cooling effect from pumping the IVC. |
| 1340 | Matt exits lab |
| 1455 | Matt enters lab |
| 1455 | Start gently pumping out IVC with Scroll Pump again this has a cooling effect, temperature stops rising, flattens out. |
| 1505 | Scroll on IVC = 0.1 mbar, and falling. So it looks like it's pumping out better now. Probably the leak has disappeared now that the IVC is warmer. |
| 1510 | Scroll on IVC = 7.4 μbar. Switch to Turbo Pump. |
| 1510 | Recovery meter = 0.08 cuft/hr. |
| 1515 | Turbo on IVC = 0.12 μbar. With Turbo valved out for 1 minute, IVC = 1.2 μbar. Continue turbo pumping. |
| 1515 | MC(Cernox) = 158.3 K |
| 1530 | Close helium recovery valve at MB exhaust. |
| 1535 | Try leak detector on IVC. With Turbo valved out for 1 minute, IVC leak signal = 510 pbarL/s. That's significantly lower than the signal seen during the LN2 blowout, maybe just because the temperature is higher. |
| 1545 | Turbo on IVC = 54 nbar. With Turbo valved out for 1 minute, IVC = 860 nbar |
| 1600 | Fill IVC slowly with N2 to around 0.46 bar on mechanical gauge. Note: digital gauge reads 190 mbar, I guess it just is not correct at highish pressures. MC temperature starts rising again |
| 1605 | Matt exits lab |