2021/04/18 (Sunday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Derek
Status Run 74 in progress. Cooling down after fixing DIB2qI short which had reappeared at around 130 K. Initial LHe transfer 2 days ago, as we approached 4K we seem to spring a leak in the IVC. (First evidence around 40 K). Can't seem to pump it out.
1115 Matt enters lab
1115 MB = 13.2%
1115 IVC = 0.4 mbar. We are not pumping on it, just monitoring it.
1115 Isolate IVC then pump out manifold with Scroll then Turbo. It pumps out just fine, falls to 10 nbar in a couple minutes. Nothing wrong with the pumps and probably the gauge is OK too.
1120 Fill LN2 cold trap bath
1120 Turn off 3He pump. Mixture still open to dump, trap1, fridge.
1120 Start pumping IVC with scroll. After 15 minutes Scroll pressure still at 0.16 mbar, not good.
1135 Valve out Scroll. After a minute IVC = 0.25 mbar.
1135 Set up fridge computer camera taped to lamp to monitor IVC pressure remotely.
1140 IVC = 0.4 mbar.
1140 Matt exits lab.
2030 Fridge temperatures are beginning to rise slightly as MB boils away
MC(Cernox)/Sorb/1K Pot = 4.30/6.2/4.94 K
2030 IVC = 0.4 mbar