2021/03/06 (Saturday)
- Lab in-person: Matt
- Virtual partner: Derek
| Status | Run 74 in progress. |
|---|---|
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/nothing → DCRC50(3)/55(1). No lead/poly shield. | |
| 1000 | MB = 69.0% |
| 1040 | Tower = 90.1 mK. More stable than the last couple days. |
| At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 100/10/400/23 e-6. A little better than last night but still bad. | |
| 1045 | Zap SQUIDs, for no particular reason |
| 1100 | Look for trigger thresholds above noise in each channel. Wound up with DCRC3qI/qO/DCRC1qI/qO = 600/60/2000/150. Note that these are all around 6 times the noise measurements. |
| 1124 | Start 31min series of +11V data (bias applied after 1 min), with diverter boards removed, with 1000 beginning randoms, trigger threshold 600/60/2000/150 on DCRC3qI/qO/DCRC1qI/qO, no in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Use sequence r74_takeandflash_dcrcbias.msl |
| Triggering looks OK in two outer channels (the lower noise channels) but it looks like the thresholds are too high to see many events in q4 and q3 | |
| 1305 | Matt enters lab |
| 1310 | Discover that the transfer line o-ring is still present - Ruiren was mistaken yesterday. |
| 1320 | Adjusting the cardboard and magazines under the DCRCs, DCRC1 shut off a couple of times, had to tap it to get it to start again. Cycled the DAQ programs and started a series to see if MIDAS was still working - and it looks like the noise is much lower now! |
| 1330 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 2.5/1.1/10/1.4 e-6. Much improved - the two worst channels have 40 times less noise than last time I checked. |
| 1338 | Start 31min series of +11V data (bias applied after 1 min), with diverter boards removed, with 1000 beginning randoms, trigger threshold 60/60/60/60 on DCRC3qI/qO/DCRC1qI/qO, no in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Use sequence r74_takeandflash_dcrcbias.msl |
| Triggering looks OK | |
| OK - we're finally seeing the 60keV line I think! And there's a hint of the time-varying impact ionization. | |
| 1345 | Replace broken gasket for cryostat-side transfer line cap with a round piece of duct tape. |
| 1355 | Tower/MC/Coldplate = 92.8/29.4/188 mK, Still = 1.47 K / 0.130 mbar / 0 mW, G1/G2 = 219/192 mbar, 1K Pot = 1.67 K / 35±2 mbar. |
| 1500 | Finish taking measurements of fridge frame and shield, for UCDenver folks designing NaI detector holder |
| 1505 | Fill LN2 cold trap bath |
| 1510 | Matt exits lab |


