| Status | End of Run 73 warmup. 1K Pot pumped out, presumably dry, and isolated. All mixture in dump. Fridge and 4He trap isolated. Recovery valve closed. N2 gas in IVC and cryostat vacuum jacket. |
| 0800 | MC(Cernox)/Sorb = 236.5/>250 K |
| 0840 | MC(Cernox) = 237.4 K. Assuming room temperature of 295 K this suggests tau_warming = 37.6 hours. That's pretty slow. Assuming dew point of 279 K (that's guessing lab humidity of 35%) fridge will reach dew point Wednesday morning. |
| 1000 | Ruiren temperature 36.6 C |
| 1420 | Ruiren enters lab |
| 1430 | Move lead from shield into LIGO lab |
| 1500 | Ruiren leaves lab |
| 1600 | MC(Cernox) = 248.4 K. Looks like tau_warming is faster now, 32.7 hours. Should reach dew point in wee hours Tuesday night. |