2020/10/25 (Sunday)
- Lab in-person: Matt
- Virtual partner: Ruiren, Derek
| Setup | Run 73. DIB 1/3/4 → Stripline 43/48/41 → pi-filters 1/2/3 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)-master; DCRC33(2) → HV control. DCRCs on. Lead/poly shield erected (4 lead walls, 2 poly walls). |
|---|---|
| 0850 | Tower thermometer wandering between 92.8 and 93.5 mK. IbIs on DCRC1pD implies detector temperature around 60.2 mK. Pattern consistent with last night. |
| 0856 | Start a 60 min series at HV = +10V, drifting 0.0225 V/s, no prebias, CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = set automatically to track temperature, Front-End Gains = (3/3/3)/(5/5/5) 1000 BOR & 5 Hz in-run randoms, trigger on DCRC1pB/D threshold = 8/8 Use RootSequencer.c, function drift_sequence(10,0.0225,3600,0,0) Use QETbias_update.c, function sequence_update_QETbiases(62,30,0,0.0885,52.04) |
|---|---|
| 0856 | Vtest = 0.2 mV shortly after drift begins. Drift starts at 10.2 V. 0.78% HV loss. |
| Tower thermometer wandered between 92.7 and 93.4 mK during series. | |
| 0957 | Vtest = 0.48 V at end of drift. Final HV = +91.5 V. 0.21% HV loss. |
| 1055 | Double-click Arduino_Create_Bridge on desktop, then go to Arduino Web Editor and select Monitor. This causes the Tower temperature to spike to 1.4 K, then fall. |
|---|---|
| 1255 | Matt Fritts body temperature = 96.4 F |
| 1340 | Matt enters lab |
| 1345 | Tower/MC/Coldplate = 93.3/31.8/199 mK, Still = 1.55 K / 0.130 mbar / 0 mW, G1/G2= 250/220 mbar, 1K Pot = 1.62 K / 21±2 mbar. |
| 1413 | Begin LHe transfer @ MB = 18.2% |
| 1420 | Swap HV leads to negative polarity |
| 1425 | Fill LN2 cold trap bath |
| 1447 | Finish LHe transfer @ MB = 93.7% |
| 1515 | MB = 93.3% |
| 1530 | Measure resistance to Backward Endstop on source mover. This reads O/L, since the source carriage is still 0.5 mm from the stop. Tried to set up the Keithley to measure this (instead of measuring V across the HV test resistor) but can't figure out how to make it stay in resistance measurement mode, keeps switching back to voltage. |
| 1700 | Matt leaves lab |
| 1955 | Do more IbIs to get detector temperature versus Tower thermometer relationship. Detector around 59.4 mK, Tower thermometer around 88 mK. |
|---|---|
| 2006 | Start a 60 min series at HV = -100V, drifting 0.0290 V/s, prebias 15% for 2min, CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = set automatically to track temperature, Front-End Gains = (3/3/3)/(5/5/5) 1000 BOR & 5 Hz in-run randoms, trigger on DCRC1pB/D threshold = 8/8 Use RootSequencer.c, function drift_sequence(100,0.0290) Use QETbias_update.c, function sequence_update_QETbiases(63,30,0,0.0337,56.41) |
| 2006 | Vtest = -0.55 mV shortly after drift begins. Drift starts at -107.6 V. 0.20% HV loss. |
| Tower thermometer wandered between 86.6 and 88.3 mK during series | |
| 2107 | Vtest = -0.86 V at end of drift. HV = -211.4 V. 0.16% HV loss. |