| 1127 | Seeing a larger leak at KF flange to elbow. ~400-700 pbar*l/s. c/w top of elbow, more like 100 pbarL/s. |
| 1145 | Slowly letting nitrogen into IVC, bringing it to 1atm. |
| 1200 | IVC up to 1atm. |
| 1210 | Open KF flange clamp at stripline elbow. Clean metal surfaces with methanol and o-ring with wipe. Add thin coat of vacuum grease to o-ring. |
| 1212 | Start slowly scroll pumping on IVC. |
| 1223 | At 0.1 mbar switch from scroll pump to turbo pump. |
| 1228 | Turbo on IVC = 190 ubar. |
| 1228 | Leak signal on IVC with turbo = 5 pbar*l/s. |
| 1229 | Leak signal with turbo valved out = 50 pbar*l/s. |
| 1230 | Tried blasting near KF flange and elbow again, no leak seen. |
| 1247 | Matt leaves lab. |
| 1750 | Matt Fritts body temperature = 97.8 F |
| 1240 | Turbo on IVC = 63 nbar. |
| 1310 | Turbo on IVC = 30 nbar. |
| 1340 | Turbo on IVC = 22 nbar. |
| 1410 | Turbo on IVC = 19 nbar. |
| 1440 | Turbo on IVC = 16 nbar. |
| 1540 | Turbo on IVC = 14 nbar. |
| 1840 | Turbo on IVC = 12 nbar. |