2020/02/29 (Saturday)

Previous Day

Setup Run 72. DIB 1/3/4 → Stripline 43/41/48 → Cable 3/2/1 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1). DCRCs on overnight. No shield.
1120 Tower/MC/Coldplate = 53.6/37.5/T.Over(NaN) mK, Still = T.Over (NaN)/ 0.110 mbar / 0 mW, G1/G2= 369/337 mbar, 1K Pot = 2.09 K / 24 mbar. MB = 67.0%
1120 Fill LN2 cold trap bath
1140 Labview appears to be misbehaving, reading NaN for several sensors and negative for Tower
1220 Changed resistance range on LakeShore from 200kohm to 63.2kohm, readings appear to be looking more normal now
1225 Coldplate = 185mK Still = 1.31K
1230 Tower/MC/Coldplate = 59.2/45.2/185 mK, Still = 1.31/ 0.110 mbar / 0 mW, G1/G2= 378/349 mbar, 1K Pot = 2.09 K / 24 mbar. MB = 67.0%
1230 Tower/MC now seems a little too warm… tried switching LakeShore resistance range back to 200kohm but no change in readings