| Setup | Run 72. DIB 1/3/4 → Stripline 43/41/48 → Cable 3/2/1 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1). DCRCs on overnight. No shield. |
| 1120 | Tower/MC/Coldplate = 53.6/37.5/T.Over(NaN) mK, Still = T.Over (NaN)/ 0.110 mbar / 0 mW, G1/G2= 369/337 mbar, 1K Pot = 2.09 K / 24 mbar. MB = 67.0% |
| 1120 | Fill LN2 cold trap bath |
| 1140 | Labview appears to be misbehaving, reading NaN for several sensors and negative for Tower |
| 1220 | Changed resistance range on LakeShore from 200kohm to 63.2kohm, readings appear to be looking more normal now |
| 1225 | Coldplate = 185mK Still = 1.31K |
| 1230 | Tower/MC/Coldplate = 59.2/45.2/185 mK, Still = 1.31/ 0.110 mbar / 0 mW, G1/G2= 378/349 mbar, 1K Pot = 2.09 K / 24 mbar. MB = 67.0% |
| 1230 | Tower/MC now seems a little too warm… tried switching LakeShore resistance range back to 200kohm but no change in readings |