2020/02/27 (Thursday)

Previous Day

Setup Run 72. DIB 1/3/4 → Stripline 43/41/48 → Cable 3/2/1 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1). DCRCs on overnight. No shield.
overnight: tower temperature falling at constant rate, 0.05 mK/hour, 1.1 mK/day
0910 Tower/MC/Coldplate = 53.5/33.1/198 mK, Still = 1.55±0.05 K/ 0.200 mbar / 0 mW, G1/G2= 331/302 mbar, 1K Pot = 2.14 K / 24 mbar. MB = 68.3%
0910 Fill LN2 cold trap bath
1830 Reconnected the Kepco HV by putting the black lead back into the -.
1915 Start a series at HV = 300→452V, rate 46mV/s, 55 minutes, with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (na/360/430)/(360/420/440) uA, Front-End Gains = (na/3/3)/(5/5/5)
1000 BOR and 5Hz in-run randoms, trigger on DCRC1pD threshold = 9
Use sequence HVdrift_250V_45mVps_pre10-2_60min.msl
1915 Tower = 52.5 mK
1929 Data starts
1935 Vtest = 4.5 mV. pulses on BatViewer look OK.
Vtest looks a lot more stable today than it did before.
2000 Vtest = 5.0 mV. pulses on BatViewer look OK.