start › cdms › k100 › run_diary › 20200221

2020/02/21 (Friday)

Previous Day

Setup Run 72. DIB 1/3/4 → Stripline 43/41/48 → Cable 3/2/1 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1). DCRCs on overnight. No shield.
overnight: tower temperature falling at constant rate, 0.16 mK/hour, 3.7 mK/day
0855 Tower/MC/Coldplate = 64.4/31.9/192 mK, Still = 1.57 K/ 0.200 mbar / 0 mW, G1/G2= 133/127 mbar, 1K Pot = 2.11 K / 26 mbar. MB= 65.2%
0900 Fill LN2 cold trap bath from 25L bomb Dewar
0935 There are good pulses in all channels except HV1pA (DCRC4pA). That channel has OK SQUID performance but the baseline seems unresponsive to changes in QET bias. Let's try a different DCRC.
0945 Swap out DCRCX(4) for DCRC50(3) on HV1.
0955 Zap SQUIDs on DCRC3
1010 DCRC3 phonon channels are railed. Try SQUID tuning on a couple channels - behavior doesn't make sense. With SQUID Bias at zero I see a trace (polarity open, source pre-amp) but when you change the SQUID bias the baseline moves a lot - just -1.1 uA and the baseline rails, should be able to set it to -70 or something.
1045 Try power-cycling DCRC3
1105 Zap SQUIDs on DCRC3. But there doesn't seem to be any response to temperature or Still pressure. So maybe it's not working.
1110 Same problem with SQUID tuning with DCRC3. No idea what's wrong.
Crap! just realized I did not have the DIB1 cable plugged into the HV box, it was plugged into the breakout box to check interchannel connections (which seem OK). So let's try this again.
1115 Power up DCRC3 again on HV1 and zap SQUIDs. This time I see fridge response.
1120 OK, now look at DCRC3 pA, pB, pC. pB and pC show pulses as before. pA is unresponsive to changes in QET bias. The simplest conclusion is that DIB1pA QET is open. Possibly one wirebond broke at the detector some time after the tower electrical checkout - during installation or during cooldown. (This TES has the most complicated wirebonding at the detector, involving a hopscotch arrangement.)
So: Side1pE is not working. That's HV1 pA.
1125 For no particular reason, switch back to DCRCX(4) on HV1. Zap DCRC4 SQUIDs.
1720 Fill LN2 cold trap bath from 25L bomb Dewar
1720 Tower = 63.5 mK
1725 Main bath = 52.7%