2019/09/25 (Wednesday)

Previous Day

Status: Started 1K Pot yesterday. Can't cool below 3 K, probably because of leak into IVC.
0920 MC(Cernox)/Tower/Sorb = 3.16/3.5/4.0 K, G1(fridge) = 307 mbar, 1K Pot = 1.91 K / 45 mbar
0920 Turbo on IVC = 250 nbar. IVC = 16 μbar with pump valved out
Leave pump valve closed to see if if creeps up.
0925 Fill LN2 cold trap
0950 IVC = 40 μbar
0950 MB = 39.6%
1015 IVC = 86 μbar
let's stop the 1K Pot and let the fridge warm back up to 4K
1020 Close 1K Pot needle valve. 1K Pot pressure falls, temperature rises. Fridge pressure starts rising. Tower temperature is not responding as quickly.
1025 1K Pot = 4.03 K / 0.36 mbar. Close 1K Pot speedivalve, leaving it essentially empty. Switch 4He pump outlet to Exhaust, close 1K Pot recovery system valve, turn off 4He pump.
1030 Tower → 4.1 K. G1(fridge) → 481 mbar.
1030 IVC = 120 μbar
start pumping out IVC with turbo pump again.
1100 Tower/1K Pot/Sorb = 4.23/4.66/5.7 K, G1(fridge) = 523 mbar
1145 Tower/1K Pot/Sorb = 4.28/4.64/5.7 K, G1(fridge) = 531 mbar
1145 Put leak detector on IVC and valve out turbo pump.
1145 IVC = 3.1e-2 mbarL/s.
1145 Start pumping mixture from fridge to dump/Trap1. Condenser & Still valves, V1, V3 open, 3He pump on, V6 open to 12.8%, V13a & V9 open. When P1 starts falling open V6 further.
1150 IVC = 3.7e-2 mbarL/s. G1/P1 (fridge) = 2.6/0.6 mbar.
1150 IVC = 4.7e-2 mbarL/s. G1/P1 (fridge) = 1.3/0.04 mbar.
1215 IVC = 5.7e-2 mbarL/s. G1/P1 (fridge) = 1.2/0.02 mbar.
1225 IVC = 6.0e-2 mbarL/s. G1/P1 (fridge) = 1.4/0.017 mbar.
so again, looks like pumping out the fridge does not cause the IVC leak signal to decrease. So there's no leak from fridge to IVC. Must be from main bath to IVC.
1235 Stop pumping out fridge, close Still and Condenser valves and V1.
Pump cold LN2 Trap 1 contents to dump.
1240 Stop leak checking IVC and start pumping it out with the turbo pump again.
1330 Check DIB3 Feedback A at breakout box : 10 kOhm. The others read about 1.2 kOhm. This channel first showed problems at 77K, where it was reading 1.4 MOhm while the others read around 10 kOhm.
1342 Let's look at some SQUIDs at 4.25 K. Connect DIB1 to DCRC1, power it up and start MIDAS programs and zap DIB1 SQUIDs.
1426 DIB1PA SQUID is not responding to any changes in SQUID bias. Not sure what's wrong with it. Other channels look fine on this DIB. Stop SQUID tuning and shut down programs and DCRC.
1434 Check DIB1 PA SQUID bias / Feedback coil = 113 Ω / 1.26kΩ at the breakout box, similar to other channels. Maybe it's a DCRC issue?
1435 Connect DIB2 to DCRC1, power it up and start MIDAS programs and zap DIB2 SQUIDs.
1440 Again DIB2PA SQUID is not responding, but other channels look fine, this is probably a DCRC issue. Maybe it can't latch the test signal to PA FB or QET?
It also looks like DCRC1 PB QET can't disconnect from the test signal. If the test signal is turned on, it shows up at the PB QET bias.
1453 Power down DCRC1. Disable DCRC1 and enable DCRC3 in ODB. Connect DCRC50 (aka DCRC3) to DIB1, start up programs, and zap DIB1 SQUIDs.
Looks like DCRC3 can't connect a test signal to any channel.
1534 Power down DCRC3. Disable DCRC3 and enable DCRC4 in ODB. Connect DCRCX (aka DCRC4) to DIB1, start up programs, and zap DIB1 SQUIDs.
All channels look good when using this DCRC. SQUID bias settings for DIB1 PB,PC,PD are similar when using DCRCs 55 and X, but the lockpoints are quite different.
Channel DCRC SQUID Bias Modulation Depth (Vpp) Lockpoint (V)
DIB1PA 55 - - -
DIB1PB 55 -68 0.127 -2.6
DIB1PC 55 -65 0.13 -3.5
DIB1PD 55 -84 0.146 -2.3
DIB1PA X -64 0.144 -5.1
DIB1PB X -68 0.134 -4
DIB1PC X -67 0.139 -4.9
DIB1PD X -85 0.149 -5
DIB2PA 55 - - -
DIB2PB 55 -83 0.144 -2.2
DIB2PC 55 -82 0.141 -3.2
DIB2PD 55 -85 0.17 -1.9
1605 Stop programs and power down DCRC.
1628 Turn on DCRC X attached to DIB2. Zap the squids
Channel DCRC SQUID Bias Modulation Depth (Vpp) Lockpoint (V)
DIB2PA X -95 0.164 -4.5
DIB2PB X -84 0.147 -3.6
DIB2PC X -91 0.178 -5.1
DIB2PD X -90 0.19 -5
1651 Power down DCRC X. Switch connection to DIB3, power on DCRC and zap squids.
Channel DCRC SQUID Bias Modulation Depth (Vpp) Lockpoint (V)
DIB3PA X -80 0.14 -4.5
DIB3PB X -65 0.148 -3.6
DIB3PC X -75 0.154 -4.3
DIB3PD X -75 0.104 -4
1710 DIB3PA was the channel that was reading a high feedback resistance at a low temperature electrical check, the channel appears to be working fine which is a good sign.
1715 Power down DCRC X.

Next Day