UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20190703

2019/07/03 (Wednesday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Three walls of the lead/poly shield are up.
0850 Tower/MC/Coldplate = 38.8/29.4/197 mK, Still = 1.36 K / 0.18 mbar / 6 mW, G1/G2 = 197.1.1/190.5 mbar, 1K Pot = 1.615 K / 19 mbar. MB = 69.1%
0930 Attempted to fill LN2 Cold trap but there was some kind of blockage.
1510 Fill LN2 cold trap from transfer dewar.
1630 Experiment with DCRC1pD trigger threshold, with the new trigger length settings. It looks like a threshold of 3 starts to dip into the noise. We'll try that.
1643 Start a sequence of 6 @ 87 min series at HV = 180→400V, rate 2.49V/min, with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/495)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun randoms, trigger on DCRC1pD threshold = 3, baseline/sample length = 20/30 (standard 5/16), phonon pre-pulse 600 (standard is 500).
Use sequence HVdrift_6X180V_2p49VPM_loopversion_prebias_10_2_dur_87min_flashDCRC_1_4.msl

Next Day