2019/06/29 (Saturday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Three walls of the lead/poly shield are up.
1332 Tower/MC/Coldplate = 38.9/29.8/198 mK, Still = 1.38 K / 0.18 mbar / 6mW, G1/G2 = 217.1/203.0 mbar, 1K Pot = 1.62 K / 20.0 mbar. MB = 62.6%
1340 Fill LN2 cold trap bath from transfer dewar.
1221 Start a sequence of 6 @ 87 min series at HV = 180→400V, rate 2.49V/min, with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/495)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun randoms, trigger on DCRC1pD threshold = 6.
Use sequence HVdrift_6X180V_2p49VPM_loopversion_prebias_10_2_dur_87min_flashDCRC_1_4.msl