Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Three walls of the lead/poly shield are up.
0925 Tower/MC/Coldplate = 40.9/30.9/201 mK, Still = 1.29 K / 0.090 mbar, G1/G2 = 150/141 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 34.5%
0930 Fill LN2 cold trap bath
1010 Remove low-pass filter from HV line.
1015 Let's try to understand the 0V noise better. We'll take SC data to get a look at the Passive Johnson Noise (PJN) and we'll also need the IbIs slope in SC and transition to see how much the PJN contributes to noise in transition. Start with IbIs slopes. We'll measure each channel with all the others still in transition.
Channel Ib (uA) Vo (V) Notes Ib/Is slope
DCRC1PB 415 -0.8778 28.8116428442194
420 -0.8759
425 -0.8696
-5 0.6597 -2.35381571640133
0 0.6041
5 0.5497
DCRC1PC 475 -0.5671 12.35495951417
480 -0.5557
485 -0.5462
-5 0.5107 -2.5134286188191
0 0.4604 Drippy noise
5 0.4077 Strong resonance at 100 kHz
DCRC1PD 445 0.0616 3.95275816581158
450 0.0949
455 0.1271
-5 0.051 -2.08814335055207
0 -0.0111
5 -0.073
DCRC4PA 485 -0.9169 8.81225781721876
490 -0.9009
495 -0.8876
-5 -0.2884 -3.65187660253175
0 -0.3043
5 -0.3536
DCRC4PB 395 -1.0005 6.45581928792262
400 -0.9787
405 -0.9605
-5 0.1943 -1.95417940344415
0 0.1278
5 0.0618 unstable, slowly rises
DCRC4PC 485 -1.035 strong oscillations
488 -1.0198 6.2730242719178
490 -1.0106
495 -0.9907
-5 0.3482 -2.22257019454358
0 0.2898
5 0.2317 unstable, slowly rises
1121 Start a 5 min SC series at HV = 0 V
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (0/0/0)/(0/0/0) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun randoms
1128 Return QET biases to standard values: DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA
1130 Set SQUID bias and lockpoint of unused channels (DCRC1PA and DCRC4pD) to 0.
1154 Begin strong neutron activation of detector by hanging PuBe source in small bucket inside NW corner of shield. (targeting 2hr4min)
Tower temperature increases 40.9→41.9 mK when source is introduced
1358 End strong neutron activation by returning PuBe source to its paraffin-filled transfer tub. Almost exactly 124 minutes of strong activation
Tower temperature returns to base after about 5 minutes.
1400 Return PuBe source to storage in PAN 443
1410 Check channels - all look good (thanks Nick). Maybe should wait till after transfer to take more data.
What trigger threshold should we use? At 6/6/6 we only got 25% write efficiency; at 10/10/10 it was 90%. Let's try 10/10/10.
1502 Started LHe transfer. MB = 27.0%
1615 Fill LN2 cold trap bath.
1618 Start an 87 min series at HV = 180→400V, rate 2.52V/min, with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 10/10/10.
Use sequence HVdrift_180V_cal0612_2p52VPM_prebias_10_2_dur_87min_flashDCRC_1_4.msl
2052 Start an 87 min series at HV = 180→400V, rate 2.52V/min, with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 10/10/10.
Use sequence HVdrift_180V_cal0612_2p52VPM_prebias_10_2_dur_87min_flashDCRC_1_4.msl
DCRC4pB was railed for this entire series