2019/05/23 (Thursday)

Previous Day

Status: 1K pot clogged at 4K, warming back up. Helium exchange gas in cryostat jacket and IVC.
0900 MC(Cernox) = 280.7 K, Vacuum jacket = 1.2 mbar
Start pumping out vacuum jacket with scroll pump.
0905 LN2 Trap 1 is room temp. Check contents: 4.8-1.4 = 3.4 mbar on G1. Pump out with 4He pump from both sides of the trap by opening both V2 and V11A.
0912 Start pumping vacuum jacket with turbo pump.
0920 Turbo on cryostat vacuum jacket = 220 nbar
0922 Put cold trap 1 in LN2 bath and top off LN2 bath.
0930 Turn off Still and MC heaters.
0935 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.4%
0938 Turbo on cryostat vacuum jacket = 74 nbar
0945 Increase V6→13.0% to restart flow
0952 Increase V6→13.1% to restart flow
0955 Turbo on cryostat vacuum jacket = 46 nbar
1100 Stop cleaning through Trap 1 after 85 min. G2 flow pressure was 570-590 mbar. Pump mixture to dump.
Removed 5.7 mbar
1105 Turbo on cryostat vacuum jacket = 30 nbar.
1122 Take cold trap 1 out to warm up.
1200 Turbo on cryostat vacuum jacket = 25 nbar. With pump valved out, cryostat vacuum jacket = 250 nbar.
Stop pumping on cryostat vacuum jacket valve.
1205 IVC = 12 mbar on digital gauge. Begin pumping out IVC with scroll pump.
1215 Switch over to pumping on IVC with turbo.
1235 LN2 Trap 1 is room temp. Check contents: 6.5-0.8 = 5.7 mbar on G1. Pump out with 4He pump from both sides of the trap by opening both V2 and V11A.
1252 Put cold trap 1 back in LN2 bath.
1253 Turbo on IVC = 30 nbar
1258 Check pressure in condenser line (includes LHe cold trap) at G1 by opening V1. 2.0-1.6 = 0.4 mbar.
Check pressure in fridge at condenser side on G1 by opening condenser valve. 2.1-2.0=0.1 mbar.
Check pressure in fridge at still side on P1 by opening still gate valve: P1=0.16 mbar.
Pump out both sides of fridge with scroll pump.
1320 G1/P1 = 0.9/0.2 mbar. P1 is still decreasing slowly.
1400 G1/P1 = 0.9/0.11 mbar. Stop pumping out fridge lines.
1404 Open dump to both sides of fridge lines via LN2 cold trap 1.
1405 With 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump.
Once P2 < 100 mbar, open 1K pot speedivalve and needle valve (5 turns) to pump out 1K pot volume.
1415 P2 < 1 mbar, close V4A. Open V5A, V2A, and the vent speedivalve to evacuate the clear hose leading to the He gas line.
Close off scroll pump. Flush MB and 1K pot (both speedivalve and needle valve open) with 320 mbar (above 1 atm) of 4He gas. Close the 1K Pot speedivalve and needle valve.
1420 For the 2nd time, with 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump.
Once P2 < 100 mbar, open 1K pot speedivalve and needle valve (5 turns) to pump out 1K pot volume.
1430 P2 < 1 mbar, close off scroll pump. Pressurize MB and 1K pot (both speedivalve and needle valve open) with 280 mbar (above 1 atm) of 4He gas. Close 1K Pot needle valve and speedivalve.
1433 Turbo on IVC = 16 nbar. IVC = 120 nbar with pump valved out.
1440 Let 5.4 mbar He exchange gas into IVC
1445 Begin LN2 transfer.
1450 MC(Cernox) = 282.8 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 616 mbar, Level resistor = 0.996 kΩ
1452 Like we did on 2019/05/18 , let's flow mix through the fridge as we cool. Open V6→12.4%. Then close V3.
Now we have the circulation pump pulling at the Still side, it's output is open to the dump via V9 and cold trap 1. Cold trap 1 is open to the condenser side.
1458 G1/G2 = 616/557 mbar and flat.
1500 MC(Cernox) = 281.8 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 615 mbar, Level resistor = 1.063 kΩ. DIB4 pD SQUID = 3.2 kΩ.
1503 Start some flow by closing V13A and opening V6 a few clicks to V6→13.1%, where G2 starts to rise a little bit. Reopen V13A.
1509 G1/G2 = 730/663 mbar and flat.
1510 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 15.5 mbar/min and G2 remains flat.
1515 MC(Cernox) = 270.6 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 730 mbar, Level resistor = 1.143 kΩ. DIB4 pD SQUID = 3.1 kΩ.
1525 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 18.3 mbar/min
1530 MC(Cernox) = 248.7 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 728 mbar, Level resistor = 1.221 kΩ. DIB4 pD SQUID = 2.7 kΩ.
1540 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 19.9 mbar/min
1545 MC(Cernox) = 221.1 K, Sorb = >249 K (max), IVC = 3.0 mbar, G1 (fridge) = 727 mbar, Level resistor = 1.302 kΩ. DIB4 pD SQUID = 2.3 kΩ but unstable.
The level resistor doesn't usually read this high so early on in the transfer. Playing with the wiring doesn't seem to change things.
1550 Take out level resistor and dip in in the LN2 cold trap bath, it reads 1.343 kΩ. Put level resistor back in main bath port.
1555 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 25.5 mbar/min
1600 MC(Cernox) = 195.6 K, Sorb = >249 K (max), IVC = 2.5 mbar, G1 (fridge) = 725 mbar, Level resistor = 1.322 kΩ. DIB4 pD SQUID = 2.0 kΩ
1610 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 31.9 mbar/min
1615 MC(Cernox) = 175.0 K, Sorb = 209.6 K, IVC = 2.2 mbar, G1 (fridge) = 723 mbar, Level resistor = 1.335 kΩ. DIB4 pD SQUID = unstable
1625 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 36.5 mbar/min
1630 MC(Cernox) = 160.7 K, Sorb = 179.8 K, IVC = 2.2 mbar, G1 (fridge) = 720 mbar, Level resistor = 1.341 kΩ. DIB4 pD SQUID = unstable
1640 Close V12A speedivalve for a minute, then re-open it. G1 decreases at 41.9 mbar/min
1645 MC(Cernox) = 148.6 K, Sorb = 158.2 K, IVC = 2.2 mbar, G1 (fridge) = 717 mbar, Level resistor = 1.343 kΩ. DIB4 pD SQUID = 1.7 kΩ
1646 Cease LN2 transfer and put ping pong ball on main bath port
1650 Close V6 and open V3 to leave the dump open to both sides of the fridge via cold trap 1.
1700 MC(Cernox) = 137.7 K, Sorb = 143.3 K, IVC = 2.2 mbar, G1 (fridge) = 607 mbar, DIB4 pD SQUID = unstable
1730 MC(Cernox) = 122.1 K, Sorb = 130.4 K, IVC = 2.2 mbar, G1 (fridge) = 608 mbar, DIB4 pD SQUID = unstable
1910 MC(Cernox) = 93.6 K, Sorb = 122.9 K, IVC = 1.6 mbar, G1 (fridge) = 610 mbar, DIB4 pD SQUID = unstable
2100 MC(Cernox) = 81.9 K, Sorb = 122.9 K, IVC = 1.6 mbar, G1 (fridge) = 612 mbar, DIB4 pD SQUID = unstable