2019/05/17 (Friday)
| Status | Blockage found, 4th attempt. Warming up. Cryostat vacuum jacket and IVC pumped out. Mix open to cold trap 2 overnight. |
|---|---|
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 |
| 0900 | MC(Cernox)/Sorb = 253.5/>249 K. |
|---|---|
| 0905 | Pump mix to dump. Removed 0.9 mbar from Trap 2 overnight passive cleaning. |
| 0933 | LN2 Trap 1 is mostly warm. Check contents: 5.1-0.8 = 4.3 mbar on G1. Pump out with 4He pump. |
| 1000 | Still trying to figure out why we keep getting blockages. Maybe we have a little leak whenever we clean? There's contradictory evidence, but maybe it could be a cold leak through the o-rings at the LN2 traps? Let's do less aggressive cleanings today. Also, we'll try wrapping those connections in cling wrap. That certainly won't stop a leak, but maybe it'll force the air to move along the cold tubing and freeze out water before reaching the seals? |
| 1015 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.4% |
|---|---|
| 1020 | Increase V6→13.1% to restart flow |
| 1115 | Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 570-870 mbar. Pump mixture to dump. Removed 4.5 mbar. |
| 1150 | LN2 Trap 2 is room temp. Check contents: 1.1-0.2 = 0.9 mbar on G1. Pump out with 4He pump. That is from the overnight passive cleaning, not very much at all… |
| 1305 | LN2 Trap 1 is room temp. Check contents: 4.5-0.0 = 4.5 mbar on G1. Pump out with 4He pump. |
| 1330 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 13.0% |
| 1345 | Increase V6→13.2% to restart flow |
| 1415 | Increase V6→13.3% to restart flow |
| 1530 | Stop cleaning through Trap 1 after 120 min. G2 flow pressure was 570-770 mbar. Pump mixture to dump. Removed 4.0 mbar. |
| 1640 | LN2 Trap 1 is room temp. Check contents: 4.0-0.0 = 4.0 mbar on G1. Pump out with 4He pump. |
| 1700 | Top off LN2 trap. |
| 1710 | Set up passive cleaning with both sides of dump and pump contents exposed to Trap 1. With 3He pump on, open V9, V4, V12A+speedi, V13A. |