Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
NaI detector: 1.1 kV (for real) → NaI B/C (dynodes terminated) → Ortec 452(Neg input, gain=2.5, shaping = 6 us, output=+10V)/ Ortec 575A(Neg input, gain=5, shaping = 3 us) → x10/x10 attenuator → DCRCX(4) PB/PA
DCRC33(2) → kepco 1kV power supply → HV input
status: Na22 -125V data overnight, tower = 29.6 mK
0910 Tower/MC/Coldplate = 29.6/30.3/187 mK, Still = 0.0 mW / 1.3±0.1 K / 0.085 mbar, G1/G2 = 116/111 mbar, 1K Pot = 1.59 K / 17 mbar, Turbo on IVC = 0.74 nbar, MB = 71.0%
0912 Fill LN2 cold trap bath.
Tuesday's IbIs worked for all channels except DCRC3pD. We'll redo that. (Note that we don't have the PuBe now. Will do it with the Na22 present.)
1305 Prep for DCRC3pD IbIs. Turn off triggering. Reset DCRC3pD Driver Gain from 2 to 1. Then take QET biases to where IbIs is max (see May 31) and adjust driver offset to near top of range. Readjust QET bias to standard transition value.
1310 Notice that in-run randoms rate is set to 20 Hz probably since Tuesday's IbIs. So the last few days of data are polluted with randoms.
1315 Dial HV up to -125V with 15%/5min prebias using change_HV_0_to_125_15-5PreBias_DCRC2_250V.msl
1339 Start IbIs sequence on DCRC3pD, from 450 to 200 step -2 uA, HV = -125 V, front-end gain 10, driver gain 1. PuBe source present. Using IbIs_loop_DCRC3pD.msl

to return to normal after IbIs:

  • set all phonon driver gains x2
  • adjust driver offsets to get baselines to the usual low-ish spot
  • turn triggering back on on DCRC3 phonon channels
  • check randoms settings in triggerfe2
1435 Dial HV from -125V to 0V using change_HV_125_to_0_DCRC2_250V.msl
1450 Flash DCRCs 1,3 for 30 seconds
1527 Start 8x2hr 22Na series at HV = -125 V with 15% prebias for 5 min, split into 10 min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-125×8-15_5PreBias-DCRC2_250V-30Flash-30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/280/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10)
NaI detector: 1.1 kV (for real) → NaI B/C (dynodes terminated) → Ortec 452(Neg input, gain=2.5, shaping = 6 us, output=+10V)/ Ortec 575A(Neg input, gain=5, shaping = 3 us) → x10/x10 attenuator → DCRCX(4) PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9
1552 Fill LN2 cold trap full.
1635 Looks like Filename in ODB was set wrong, IbIs script should have reset it:
ODBSET “/Logger/Channels/0/Settings/Filename” .RUN%05d_DUMP%04d.mid.gz
Reset it manually just now, hopefully data will now continue correctly. Currently on 6th series, presumably 7th series and onward will be ok