2018/10/03 (Wednesday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
NaI detector: 1.1 kV (for real) → NaI B/C (dynodes terminated) → Ortec 452(Neg input, gain=2.5, shaping = 6 us, output=+10V)/ Ortec 575A(Neg input, gain=5, shaping = 3 us) → x10/x10 attenuator → DCRCX(4) PB/PA
DCRC33(2) → kepco 1kV power supply → HV input
status: PuBe -125V data overnight, tower = 29.5 mK
0926 Tower/MC/Coldplate = 29.5/30.3/183 mK, Still = 0.0 mW / 1.4±0.1 K / 0.090 mbar, G1/G2 = 137/142 mbar, 1K Pot = 1.60 K / 17 mbar, Turbo on IVC = 0.74 nbar, MB = 69.1%
0930 Fill LN2 cold trap bath.
1020 After 10th 10-minute series in current data block, remove PuBe source from lab and take back to Kevin Booth in Tate.
1625 Remove 22Na source from safe and place inside shield at bottom of NW corner.
1630 Fill LN2 cold trap full.
1727 Start 8x2hr 22Na series at HV = -125 V with 15% prebias for 5 min, split into 10 min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-125×8-15_5PreBias-DCRC2_250V-30Flash-30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/280/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10)
NaI detector: 1.1 kV (for real) → NaI B/C (dynodes terminated) → Ortec 452(Neg input, gain=2.5, shaping = 6 us, output=+10V)/ Ortec 575A(Neg input, gain=5, shaping = 3 us) → x10/x10 attenuator → DCRCX(4) PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9