2018/09/23 (Sunday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4) DCRC33(2) → kepco 1kV power supply → HV input |
|
| status: | PuBe -125V data overnight, tower = 29.5 mK |
| 1755 | Tower/MC/Coldplate = 29.5/32.0/170 mK, Still = 0.0 mW / 1.1±0.1 K / 0.095 mbar, G1/G2 = 127/123 mbar, 1K Pot = 1.60 K / 18 mbar, Turbo on IVC = 0.86 nbar, MB = 61.5% |
| 1759 | Fill LN2 cold trap. |
| 2248 | Start 8x2hr PuBe series at HV = -125 V with 15% prebias for 5 min, split into 10 min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-125×8-15_5PreBias-DCRC2_250V-30Flash-30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10) NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Neg) / Ortec 575A(gain=10, shaping = 3us, Neg) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9 Damn, DCRC3pC railed throughout |