UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180918

2018/09/18 (Tuesday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took bg data overnight at -50V. Long dataset without flashing.
0830 Tower/MC/Coldplate = 29.6/28.9/181 mK, Still = 1.0 mW / 1.4±0.1 K / 0.080 mbar, G1/G2 = 128/124 mbar, 1K Pot = 1.59 K / 17 mbar, Turbo on IVC = 0.86 nbar, MB = 37.5%
0835 Fill LN2 cold trap bath.
0835 Dial HV down to 0V using change_HV_50_to_0_DCRC2.msl
1110 Flash DCRCs 1,3 for 30 seconds
1505 Start LHe transfer (MB = 28.0%)
1607 End LHe transfer (MB = 90.6%)
1714 Fill LN2 cold trap full.
1733 Set HV = 50V using change_HV_0_to_50_preBias_15_5_DCRC2.msl
1741 Start 14 hr, 50 V, hold time dataset split into 10 min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence timed_subrun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10)
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9