2018/09/16 (Sunday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4) DCRC33(2) → kepco 1kV power supply → HV input |
|
| status: | Took 22Na data overnight at -80V. |
| 0940 | Sequence is still going but it looks like data is not being copied from /livedata to /byseries. Looking at Programs, Logger is on but I don't see any “lazy logger”, I think that's supposed to be running too. |
cdms2:bin cdmssoft$ps -u cdmssoft
PID TTY TIME CMD
532301 ? 10:52:46 mhttpd
532303 ? 00:01:04 mserver
638457 ? 00:02:58 dcrc_driver.exe
638462 ? 22:17:53 towerfe3.exe
638464 ? 1-06:15:46 triggerfe2_mod.
648833 ? 08:24:45 mlogger
736583 pts/6 00:00:00 bash
736771 pts/6 00:00:00 ps
3387236 pts/0 00:00:00 bash
| 0945 | Restart MIDAS. Now lazylogger is seen when you do a ps (see below). And Programs shows Lazy_0 running. Start DAQ programs. Sequence starts up where it left off. Data is being copied to /byseries. |
|---|
cdms2:bin cdmssoft$ps -u cdmssoft
PID TTY TIME CMD
736583 pts/6 00:00:00 bash
736828 ? 00:00:00 mhttpd
736830 ? 00:00:00 mserver
736832 ? 00:00:00 mlogger
736834 ? 00:00:00 lazylogger
736837 pts/6 00:00:00 ps
3387236 pts/0 00:00:00 bash
| 1258 | Tower/MC/Coldplate = 29.6/28.9/184 mK, Still = 1.0 mW / 1.5±0.1 K / 0.080 mbar, G1/G2 = 106/102 mbar, 1K Pot = 1.60 K / 18 mbar, Turbo on IVC = 1.0 nbar, MB = 28.2% |
|---|---|
| 1305 | Fill LN2 cold trap bath. |
| 1312 | Stop part way through sequence during series 07180916_1257. Turn down bias and flash detector. |
| 1320 | Commence LHe transfer, MB = 28.2% |
| 1415 | Cease LHe transfer, MB = 94.4%. Tower = 30.3 mK and settling after flash and transfer. |
| 1421 | Start 10 x 2hr Na datasets at -80V split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-80×10-15_5PreBias-30Flash-30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10) NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9 |