2018/09/02 (Sunday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4) DCRC33(2) → kepco 1kV power supply → HV input |
|
| status: | Took PuBe data overnight at -80V. |
| 1305 | Start LHe transfer (MB = 34.1%) |
| 1310 | Tower/MC/Coldplate = 30.9/29.5/185 mK, Still = 1.4±0.1 K / 0.070 mbar, G1/G2 = 123/119 mbar, 1K Pot = 1.60 K / 20 mbar, Turbo on IVC = 1.0 nbar. |
| 1318 | Fill LN2 cold trap bath. |
| 1350 | Sequence was running its 30-min waiting period after the final flash. |
| 1352 | Ended sequence. |
| 1403 | End LHe transfer (MB = 90.9%) |
| 1406 | Start 10 x 2hr PuBe datasets at -80V split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-80×10-15_5PreBias-30Flash-30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9 |