start › cdms › k100 › run_diary › 20180831

2018/08/31 (Friday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took PuBe data overnight at -30V, -50V, -40V.
1105 Tower/MC/Coldplate = 30.3/29.3/184 mK, Still = 1.4±0.1 K / 0.070 mbar, G1/G2 = 102/102 mbar, 1K Pot = 1.60 K / 19 mbar, Turbo on IVC = 1.0 nbar, MB = 34.2%
1110 Fill LN2 cold trap bath.
1205 DCRC1pB railed around 20 minutes into the last 2-hour dataset. Adjust Driver Offset (this is at 0V, so not certain will work at bias, we'll see).
Last night's bias scan showed progressively improved noise (in keVee sigma) as bias increased -30V to -50V. Let's try -70V.
1207 Start 1 x 2hr -70 V background dataset split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-70-15_5PreBias-30Flash-30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1447 Start LHe transfer (MB = 28.6%)
1543 End LHe transfer (MB = 95.6%)
1619 Fill LN2 cold trap full.
1707 Start 6 x 2hr background datasets, first at -80V, the rest at -70V split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-80_70x5-15_5PreBias-30Flash-30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1720 Data taking is underway at -80V. Vbias = -80.017, Vtest = -23.11 → Vdetector = -70.8V, only 88%. Open the LN2 vent valve a little more to increase the ebox purge.